EasyManuals Logo

Tektronix TDS1002 Programmer's Manual

Tektronix TDS1002
250 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #214 background imageLoading...
Page #214 background image
Status and Events
3−16
TDS200/1000/2000 Series Oscilloscope Programmer Manual
This technique requires less bus traffic than did looping on BUSY?.
The program can now do different tasks such as talk to other
devices. The SRQ, when it comes, interrupts those tasks and
returns control to this task.
/* Take peak-to-peak measurement on acquired data */
MEASUREMENT:IMMED:VALUE?
This technique is more efficient but requires more sophisticated
programming.
Using the *OPC? Query (the Most Simple Approach)
*OPC? places a 1 in the Output Queue once an operation is
complete. A timeout could occur if you try to read the output queue
before there is any data in it.
The same command sequence using *OPC? for synchronization
follows:
/* Set up single−sequence acquisition */
SELECT:CH1 ON
ACQUIRE:MODE SAMPLE
ACQUIRE:STOPAFTER SEQUENCE
/* Acquire waveform data */
ACQUIRE:STATE ON
/* Set up the measurement parameters */
MEASUREMENT:IMMED:TYPE PK2PK
MEASUREMENT:IMMED:SOURCE CH1
/* Wait until the acquisition is complete before
taking the measurement */
*OPC?

Table of Contents

Other manuals for Tektronix TDS1002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS1002 and is the answer not in the manual?

Tektronix TDS1002 Specifications

General IconGeneral
Bandwidth60 MHz
Channels2
Sample Rate1 GS/s
Record Length2.5k points
Vertical Sensitivity2 mV/div to 5 V/div
Vertical Resolution8 bits
Display Resolution320 x 240 pixels
Display5.7 inch monochrome LCD
Trigger ModesVideo, Edge
ConnectivityUSB 2.0
Weight2.0 kg
Input Impedance1 MΩ
Power Supply100 to 240 VAC, 45 to 440 Hz