Command Descriptions
TDS200/1000/2000 Series Oscilloscope Programmer Manual
2−55
CH<x>:PRObe?
1 sets the specified oscilloscope channel to 1X attenuation.
10 sets the specified oscilloscope channel to 10X attenuation.
100 sets the specified oscilloscope channel to 100X attenuation.
1000 sets the specified oscilloscope channel to 1000X attenuation.
<NR1>
CH2:PRObe 1000
sets channel 2 to x1000 attenuation.
CH1:PRObe?
might return CH1:PROBE 10
CH<x>:SCAle
Sets or queries the vertical gain of the specified oscilloscope
channel. This command is equivalent to adjusting the front-panel
VOLTS/DIV knob.
Vertical
CH<x>:SCAle <NR3>
CH<x>:SCAle?
<NR3> is the gain, in volts per division. The range is 5 V/div to
2 mV/div when using a 1X probe.
Arguments
Returns
Examples
Group
Syntax
Arguments

Table of Contents

Other manuals for Tektronix TDS1002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS1002 and is the answer not in the manual?

Tektronix TDS1002 Specifications

General IconGeneral
Bandwidth60 MHz
Channels2
Sample Rate1 GS/s
Record Length2.5k points
Vertical Sensitivity2 mV/div to 5 V/div
Vertical Resolution8 bits
Display Resolution320 x 240 pixels
Display5.7 inch monochrome LCD
Trigger ModesVideo, Edge
ConnectivityUSB 2.0
Weight2.0 kg
Input Impedance1 MΩ
Power Supply100 to 240 VAC, 45 to 440 Hz