Command Descriptions
TDS200/1000/2000 Series Oscilloscope Programmer Manual
2−63
NOTE . If Trigger View is active, this query returns 9.9E37 and
generates event 221 (Settings conflict).
Cursor
CURSor:VBArs:DELTa?
<NR3>
CURSor:VBArs:DELTa?
might return 8.92E–1, indicating that the time difference between
the vertical bar cursors is 0.892 seconds.
CURSor:VBArs:POSITION<x>
Positions a vertical bar cursor. The unit is specified by the
CURSor:VBArs:UNIts command, and can be in units of seconds or
frequency (Hertz). If the cursor source is an FFT math waveform,
CURSor:VBArs:POSITION is always in Hertz, regardless of the
value set by CURSor:VBArs:UNIts.
NOTE. If Trigger View is active, the query form returns 9.9E37 and
generates event 221 (Settings conflict).
Cursor
CURSor:VBArs:POSITION<x> <NR3>
Group
Syntax
Returns
Examples
Group
Syntax

Table of Contents

Other manuals for Tektronix TDS1002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS1002 and is the answer not in the manual?

Tektronix TDS1002 Specifications

General IconGeneral
Bandwidth60 MHz
Channels2
Sample Rate1 GS/s
Record Length2.5k points
Vertical Sensitivity2 mV/div to 5 V/div
Vertical Resolution8 bits
Display Resolution320 x 240 pixels
Display5.7 inch monochrome LCD
Trigger ModesVideo, Edge
ConnectivityUSB 2.0
Weight2.0 kg
Input Impedance1 MΩ
Power Supply100 to 240 VAC, 45 to 440 Hz