Command Descriptions
2−66
TDS200/1000/2000 Series Oscilloscope Programmer Manual
Refer to Waveform Commands on page 2-27 for a description of the
waveform transfer process.
Waveform
CURVe { <Block> | <asc curve> }
CURVe?
<Block> is the waveform data in binary format. The waveform is
formatted as: #<x><yyy><data> where <x> is the number of
characters in <yyy>. For example, if <yyy> = 500, then <x> = 3,
where <yyy> is the number of bytes to transfer. Refer to Block
Arguments on page 2-12 for more information.
If width is 1, then all bytes on the bus are single data points. If width
is 2, then all bytes on the bus are 2-byte pairs. Use the DATa:WIDth
command to set the width. <data> is the curve data.
<asc curve> is the waveform data in ASCII format. The format for
ASCII data is <NR1>[,<NR1>...] where each <NR1> represents a
data point.
SELect:CH1 ON
DATa:ENCdg ASCIi
DATa:SOUrce CH1
CURVe?
might return the ASCII data
:CURVE13,6,3,2,–1,–9,–14,–19,–29,–35,–67,–1,–78,–62,
–50,–31,–27,–7,0,12,29,39,45,43,41,47,41,38,33,26
DATa, WFMPre
Group
Syntax
Arguments
Examples
Related Commands

Table of Contents

Other manuals for Tektronix TDS1002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS1002 and is the answer not in the manual?

Tektronix TDS1002 Specifications

General IconGeneral
Bandwidth60 MHz
Channels2
Sample Rate1 GS/s
Record Length2.5k points
Vertical Sensitivity2 mV/div to 5 V/div
Vertical Resolution8 bits
Display Resolution320 x 240 pixels
Display5.7 inch monochrome LCD
Trigger ModesVideo, Edge
ConnectivityUSB 2.0
Weight2.0 kg
Input Impedance1 MΩ
Power Supply100 to 240 VAC, 45 to 440 Hz