Math FFT
120
TDS1000/2000-Series Digital Oscilloscope User Manual
Selecting an FFT Window
Windows reduce spectral leakage in the FFT spectrum. The FFT
assumes that the YT wave form re peats forever. Wi th an integral
number of cycles (1, 2, 3, ...), the YT waveform starts and ends at the
same amplitude and there are no discontinuities in the signal shape.
A non-integral number of cycles in the YT waveform causes the
signal start and end points to be at different amplitudes. The
transitions between the start and end points cause discontinuities in
the signal that introduce high-freque ncy transients.
Time-domain
(YT) waveform
Without windowing
FFT
Discontinuities
Center 2048
data points

Table of Contents

Other manuals for Tektronix TDS2024

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS2024 and is the answer not in the manual?

Tektronix TDS2024 Specifications

General IconGeneral
Channels4
Bandwidth200 MHz
Sample Rate2 GS/s
Record Length2.5k points
Vertical Resolution8 bits
Vertical Sensitivity2 mV/div to 5 V/div
Timebase Range5 ns/div to 50 s/div
Display Resolution320 x 240 pixels
Math FunctionsAdd, Subtract, Multiply, Divide, FFT
Input CouplingAC, DC, GND
Trigger ModesAuto, Normal, Single
Trigger TypesEdge, Video, Pulse
ConnectivityUSB
Input Impedance1 MΩ
Maximum Input Voltage300 V
InterfaceUSB
Power Supply100 to 240 V
Dimensions326 mm x 158 mm x 124 mm