Application Examples
54
TDS1000/2000-Series Digital Oscilloscope User Manual
Analyzing S ignal Detail
You have a noisy signal displayed on the oscilloscope and you need
to know more about it. You suspect that the signal contains much
more detail than you can now see in the display.
Looking at a Noisy Si gnal
The signal appears noisy and you suspect that noise is causing
problems in your c ircuit. To better analyze the noise, follow these
steps:
1. Push the ACQUIRE button to see the Acquire Menu.
2. Push the Peak Detect option button.
3. If necessary, push the DISPLAY button to see the Display Menu.
Use the Contrast Increase and Contrast Decrease option
buttons to adjust the contrast to see the noise more easily.
Peak detect emphasizes noise spikes and glitches in your signal,
especially when the time base is set to a slow setting.

Table of Contents

Other manuals for Tektronix TDS2024

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS2024 and is the answer not in the manual?

Tektronix TDS2024 Specifications

General IconGeneral
Channels4
Bandwidth200 MHz
Sample Rate2 GS/s
Record Length2.5k points
Vertical Resolution8 bits
Vertical Sensitivity2 mV/div to 5 V/div
Timebase Range5 ns/div to 50 s/div
Display Resolution320 x 240 pixels
Math FunctionsAdd, Subtract, Multiply, Divide, FFT
Input CouplingAC, DC, GND
Trigger ModesAuto, Normal, Single
Trigger TypesEdge, Video, Pulse
ConnectivityUSB
Input Impedance1 MΩ
Maximum Input Voltage300 V
InterfaceUSB
Power Supply100 to 240 V
Dimensions326 mm x 158 mm x 124 mm