Getting Started
TDS1000/2000-Series Digital Oscilloscope User Manual
9
Probe Attenuation Setting
Probes are available with various attenuation factors which affect the
vertical scale of t he signa l. The Probe Check function verifies that
the Probe attenuation option matches the attenuation of the probe.
As an alternative method to Probe Che ck, you can push a vertical
menu button (such as the CH 1 MENU but ton), and select the Probe
option that m atches the attenuation factor of your probe.
NOTE. The default setting for the Probe option is 10X.
Be sure that the Attenuation switch on the P2200 probe matches the
Probe option in the oscilloscope. Switch settings are 1X and 10X.
Attenuation switch
NOTE. When the Attenuati on switch is set to 1X, the P 2200 probe
limits the bandwidth of the oscilloscope to 7 MHz. To use the full
bandwidth of the oscilloscope, be sure to set the switch to 10X.

Table of Contents

Other manuals for Tektronix TDS2024

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS2024 and is the answer not in the manual?

Tektronix TDS2024 Specifications

General IconGeneral
Channels4
Bandwidth200 MHz
Sample Rate2 GS/s
Record Length2.5k points
Vertical Resolution8 bits
Vertical Sensitivity2 mV/div to 5 V/div
Timebase Range5 ns/div to 50 s/div
Display Resolution320 x 240 pixels
Math FunctionsAdd, Subtract, Multiply, Divide, FFT
Input CouplingAC, DC, GND
Trigger ModesAuto, Normal, Single
Trigger TypesEdge, Video, Pulse
ConnectivityUSB
Input Impedance1 MΩ
Maximum Input Voltage300 V
InterfaceUSB
Power Supply100 to 240 V
Dimensions326 mm x 158 mm x 124 mm