EasyManuals Logo
Home>Tektronix>Test Equipment>TDS2024

Tektronix TDS2024 User Manual

Tektronix TDS2024
207 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #28 background imageLoading...
Page #28 background image
Getting Started
TDS1000/2000-Series Digital Oscilloscope User Manual
9
Probe Attenuation Setting
Probes are available with various attenuation factors which affect the
vertical scale of t he signa l. The Probe Check function verifies that
the Probe attenuation option matches the attenuation of the probe.
As an alternative method to Probe Che ck, you can push a vertical
menu button (such as the CH 1 MENU but ton), and select the Probe
option that m atches the attenuation factor of your probe.
NOTE. The default setting for the Probe option is 10X.
Be sure that the Attenuation switch on the P2200 probe matches the
Probe option in the oscilloscope. Switch settings are 1X and 10X.
Attenuation switch
NOTE. When the Attenuati on switch is set to 1X, the P 2200 probe
limits the bandwidth of the oscilloscope to 7 MHz. To use the full
bandwidth of the oscilloscope, be sure to set the switch to 10X.

Table of Contents

Other manuals for Tektronix TDS2024

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS2024 and is the answer not in the manual?

Tektronix TDS2024 Specifications

General IconGeneral
BrandTektronix
ModelTDS2024
CategoryTest Equipment
LanguageEnglish

Related product manuals