Acquisition Controls
TDS3000B Serie s User Manual
3-3
Single Sequence Button
Push the SINGLE SEQ button to execute a single-shot acquisition.
The function of the SINGLE SEQ button depends on the acquisition
mode.
Acquisition mode SINGLE SEQ function
Sample or Peak Detect One acquisition of each displayed
channel is acquired, concurrently
Envelope N or Average N N acquisitions of each displayed
channel are acquired (N is user
adjustable with the general purpose
knob)
When you push the SINGLE SEQ button, the oscilloscope does the
following things:
H The trigger mode is set to Normal
H The trigger system arms and the light next to the SINGL E SEQ
button t urns on
After the single-sequence acquisition is complete, acquisition stops
and the light next to the SINGLE SEQ button turns off.
Push the SINGLE SEQ button again to acquire a new sequence, or
push the RUN/STOP button to restart continuous acqui sition.