AFA4000/E VAV Airflow Controller / p.134
Menu item
Option
Instructions
System Profile
Select the required system profile:
Bench type
Walk in
DDA
Keypad Settings
Audible Alarm
Obstruction
Choose the setting:
On
Off
Audible Alarm Sash Fault
Choose the setting:
On
Off
Audible Alarm Sash Lock
Choose the setting:
On
Off
Sash settings
1. Enable / disable
Choose the setting to permanently enable
or disable the S1 drive via the menu or
keypad:
On (via menu)
Off
User (via keypad)
2. Sensors
Tilt Switch
Light Curtain
Light Curtain Test
Auto Clear Obst Alarm
Personnel Sensor
Personnel Power Cycle
Sash String Type
Back
Choose the setting:
Disabled / Enabled
Disabled / NC Contact / NO Contact
Disabled / Enabled
Yes / No
Disabled / NC Contact / NO Contact
Yes / No
SP1-50 / SG1-120