4.2 Auto Sash control options
To open / close the fume hood sash the Auto Sash Controller is fitted with:
A Tiptronic open/close control via the Sash Position Sensor.
A Sash Low micro-switch
A Sash Low proximity switch
4.2.1 Auto Sash Controller with Tiptronic open / close control
Tiptronic Open / Close control is the default control setup. It functions via a Sash Position Sensor
(sprung potentiometer) with a draw wire (1.2m long) attached to either the sash or the
counterweight. The Auto Sash Controller can then be calibrated to give three methods of
operation:
Method A - Close only
The bottom position is calibrated.
The sash automatically drives to the bottom position,
when the user is not present.
Press the pushbutton to drive the sash to the bottom
position, when the user is present.
The pushbutton is disabled.
Method B - 2 Positions
The bottom and mid-positions are calibrated.
The sash automatically drives to the bottom position,
when the user is not present.
Press the pushbutton to drive the sash to the bottom
position, when the user is present.
Press the pushbutton to drive the sash to the mid-
position, when the user is present.
When the Tiptronic feature is enabled, manually tip the
sash upwards and the sash will drive to the mid-position.
Manually tip the sash downwards and the sash will drive
to the bottom position.
When the Auto Open feature is enabled, the sash
automatically drives to the mid-position when the user is
present.
There is a time delay following user detection.