AFA4000/E VAV Airflow Controller / p.139
Menu item
Option
Instructions
BMS Output Settings
Alarm Sounder
Enable / Disable the remote alarm
sounder.
Analog Volts
Select the required analogue voltage
output for sash position indication:
0 10V
1 10V
0 5V
1 5V
Analogue Current
Select the required output current for
sash position indication:
0 20mA
4 20mA
Relays 1 to 4:
RLY1, RLY2, RLY3, RLY4
Select the required settings for the relay:
Always Open
Always Closed
Open on Sash Alarm
Close on Sash Alarm
Open on Obstruction
Close on Obstruction
Open on Personnel
Close on Personnel
Open on Sash Open
Close on Sash Open
Open on Sash Locked
Close on Sash Locked
Exit and Abandon Changes
Return to Main menu without saving
Exit and Save Changes
Return to Main menu saving all changes