6 WiFi data logger
6.1 Technical data
6.1.1 Measurement-specific data
WiFi data
logger
T1
T2
T3
H1
H2
Order
number
0572 2001 0572 2002 0572 2003 0572 2004 0572 2005
Sensor type 1x NTC
internal
2x NTC or
door contact
external
2x thermocouple
type K/J/T
external
1x NTC/humidity
internal
1x
NTC/humidity
external
Measuring
range
-30 to
+50 °C
-50 to
+150 °C
K: -195 to
+1350 °C
J: -100 to +750 °C
T: -200 to +400 °C
-30 to +50 °C
0 to 100% RH
3
Probe-specific
Accuracy
±1 digit
± 0.5 °C ± 0.5 °C ± (0.5 °C + 0.5%
of reading)
± 0.5 °C
± 2% RH
(at 25 °C, 2 to
98 RH)
Probe-specific
Resolution 0.1 °C 0.1 °C 0.1 °C
(-200 to +999.9 °C
)
1 °C (remaining
meas. range)
0.1 °C
0.1% RH
0.1 °C
0.1% RH
WiFi data
logger
T1
T2
T3
H1
H2
Order
number
0572 2031 0572 2032 0572 2033 0572 2034 0572 2035
Sensor type 1x NTC
internal
2x NTC or
door
contact
external
2x
thermocoupl
e type K/J/T
external
1x
NTC/humidit
y internal
1x
NTC/humidity
external
3
Not for condensing atmosphere. For continuous application in high humidity
(> 80% RH at ≤ 30 °C for > 12 h, > 60% RH at > 30 °C for > 12 h), please
contact us (www.testo-international.com).

Other manuals for TESTO Saveris 2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TESTO Saveris 2 and is the answer not in the manual?

TESTO Saveris 2 Specifications

General IconGeneral
Protection ClassIP54
Measurement ParametersTemperature, humidity, pressure
ConnectivityWiFi
Data StorageInternal memory
Power SupplyReplaceable batteries
Operating Temperature-30 to +50 °C
Battery Life24 months
Humidity Range0 to 100% RH
Alarmsemail/SMS notification
SoftwareTesto Cloud software
CertificationsEN 12830

Related product manuals