Display Cartridge Cathode buffer
Yellow OK for use, but:
• 1–25 injections remain for the capillary
array,
or
• It will exceed the manufacturer's expiry
date within 2 weeks,
or
• It will exceed the maximum number of
days allowed on the instrument within
2 weeks
[1]
OK for use, but:
• 1–25 injections remain for the cathode
buffer,
or
• It will exceed the manufacturer's expiry
date within 2 days
or
• It will exceed the maximum number of
days allowed on the instrument (14 days)
within 2 days
Red The cartridge has expired:
• No injections remain for the capillary
array,
or
• It has exceeded the manufacturer's expiry
date,
or
• It has exceeded the maximum number of
days allowed on the instrument
[1]
The buffer has expired:
• No injections remain for the cathode
buffer,
or
• It has exceeded the manufacturer's expiry
date
or
• It has been installed on the instrument for
>14 days
[1]
The maximum number of days allowed on the instrument is 120 days for the SeqStudio
™
Genetic Analyzer Cartridge (Cat. No. A33671) and
180 days for the SeqStudio
™
Genetic Analyzer Cartridge v2 (Cat. No. A41331). For more information, see
SeqStudio
™
Genetic Analyzer
Instrument and Software User Guide
.
See SeqStudio
™
Genetic Analyzer Instrument and Software User Guide if a new
Cathode Buer Container is required.
See SeqStudio
™
Genetic Analyzer Instrument and Software User Guide if a new
cartridge is required.
3.
Touch Close, then touch .
The Eject plate command is disabled for a few minutes after you insert a cartridge. If
you are loading the CBC, sample plate, and cartridge at the same time, you can save
time by loading the CBC and sample plate before you insert the cartridge.
For information on loading the individual components, see:
• “Prepare and load the plate or tubes“ on page 14
• SeqStudio
™
Genetic Analyzer Instrument and Software User Guide
In the home screen:
Load the CBC, the
sample plate, and
the cartridge
Chapter 2 Prepare the samples and the instrument
Prepare the instrument