EasyManua.ls Logo

Thermo Scientific Evolution One Series - Cleaning the Instrument; Removing and Installing the Detector Module

Thermo Scientific Evolution One Series
42 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2
Evolution One Series Spectrophotometers
Maintenance
Thermo Scientific Evolution One Series Spectrophotometer User Guide 25
Cleaning the Instrument
Clean the exterior of the instrument and the sample and reference compartment interior
periodically. These compartments have a port to allow small amounts of liquid to drain from
the instrument.
To clean the instrument exterior and the sample or reference compartment interior
1. Use a lint-free cloth dampened with a weak solution of detergent and water to wipe the
surface as necessary.
2. Repeat using a cloth dampened with plain water.
3. Dry the surface with another cloth.
Related Topics
Maintenance
Removing and Installing the Detector Module
Some accessories have a dedicated detector and require removing the instrument detector
module before they can be installed.
Avoid UV radiation hazard in the detector compartment. Protect eyes and skin
from exposure.
To remove the detector module
1. Turn off the spectrophotometer main power switch (on the back panel).
2. Remove the detector access cover. See Detector Access Cover.
3. Loosen the thumbscrew that attaches the detector module to the instrument, and
carefully lift the module straight up.
NOTICE Do not allow moisture to leak into the instrument interior.
CAUTION Safety glasses with side shields or goggles with solid side pieces are the only
equipment that provides adequate eye protection against direct and reflected UV light.
NOTICE Avoid touching the connector pins on the bottom of the detector module.

Related product manuals