40 | Appendix A: Warning / Caution Summary Ultra Low Temperature Freezers
Appendix A: Warning / Caution
Summary
Warning/ Caution Summary
Warning/
Caution
Tone Heart Remote Ring-back Message Details
Bus Battery
Disconnected
Warning
High Red Yes Yes
Header: BUS Battery Disconnected Warning
Details: BUS Battery Disconnected Warning
active. Bus will be unable operate during a
power failure. May take up to 4 hrs. to clear
even with a new battery.
Ticker: BUS Battery Disconnected.
Bus Battery
Disconnected
Warning Recovered
None Yellow None None
Header: BUS Battery Disconnected Warning
Recovered
Details: BUS Battery Disconnected Warning
has recovered. BUS battery reconnected.
Select check box and press Acknowledge to
clear this notification.
Ticker: None
BUS Battery Low
Caution
None Yellow None None
Header: BUS Battery Low Caution
Details: BUS battery is low. Recommend
replacing battery. Contact customer service
for replacement battery. May take up to 4 hrs.
to clear even with a new battery.
Ticker: None
BUS Battery PM
Expiry Caution
None Yellow None None
Header: BUS Battery PM Expiry Caution
Details: Recommend replacing the BUS
battery. Reset battery expiration time using
the status screen to clear this notification.
Ticker: None
BUS Lost
Communication
Warning
High Red Yes Yes
Header: BUS Lost Communication Warning
Details: BUS Lost Communication Warning
active. A BUS communication error has
occurred. Contact customer service.
Ticker: BUS Lost Communication
BUS Lost
Communication
Warning Recovered
None Yellow None None
Message Header: BUS Lost Communication
Warning Recovered
Message Details: A BUS communication
error has recovered. Select check box and
press Acknowledge to clear this notification.
Ticker Message: None

Table of Contents

Other manuals for Thermo Scientific TSX Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo Scientific TSX Series and is the answer not in the manual?

Thermo Scientific TSX Series Specifications

General IconGeneral
TypeLaboratory Refrigerator
Cooling TypeForced air cooling
Defrost TypeAutomatic
Door TypeSolid or glass door options
Controller TypeMicroprocessor
Certifications/ComplianceUL, CE
Energy ConsumptionVaries by model; refer to product specification sheet
Voltage115V
Exterior Dimensions (H x W x D)Varies by model; refer to product specification sheet
Interior Dimensions (H x W x D)Varies by model; refer to product specification sheet
WeightVaries by model; refer to product specification sheet
ShelvesAdjustable shelves
Energy EfficiencyEnergy Star certified models available