Ultra Low Temperature Freezers Appendix A: Warning / Caution Summary | 45
Power Failure
Warning
High Red Yes Yes
Header: Power Failure Warning
Details: Power Failure Warning active. Unit in
power failure; display operating on battery
power. Check unit plug, unit circuit breaker in
the ON position, and supply voltage.
Ticker: Power Failure
Power Failure
Warning Recovered
None Yellow None None
Header: Power Failure Warning Recovered
Details: Power has been restored. Select
check box and press Acknowledge to clear
this notification.
Ticker: None
Stage 1 Failure
Warning
High Red Yes Yes
Header: Stage 1 Failure Warning
Details: Stage 1 Failure Warning active.
System may be unable to maintain cabinet
setpoint due to loss of communication, or
compressor or inverter failure. Contact
customer service. [Error code:XX.]
Ticker: Stage 1 Failure
Stage 1 Failure
Warning Recovered
None Yellow None None
Header: Stage 1 Failure Warning Recovered
Details: Stage 1 Failure Warning has
recovered. Select check box and press
Acknowledge to clear this notification.
Ticker: None
Stage 2 Failure
Warning
High Red Yes Yes
Header: Stage 2 Failure Warning
Details: Stage 2 Failure Warning active.
System may be unable to maintain cabinet
setpoint due to loss of communication, or
compressor or inverter failure. Contact
customer service. [Error code:XX.]
Ticker: Stage 2 Failure
Stage 2 Failure
Warning Recovered
None Yellow None None
Header: Stage 2 Failure Warning Recovered
Details: Stage 2 Failure Warning has
recovered. Select check box and press
Acknowledge to clear this notification.
Ticker: None
System Battery
Disconnected
Warning
High Red Yes Yes
Header: System Battery Disconnected
Warning
Details: System Battery Disconnected
Warning active. Unit will not be able to warm
the user during power failure. May take up to
4 hrs. to clear even with a new battery.
Ticker: System Battery Disconnected
Warning/ Caution Summary
Warning/
Caution
Tone Heart Remote Ring-back Message Details

Table of Contents

Other manuals for Thermo Scientific TSX Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo Scientific TSX Series and is the answer not in the manual?

Thermo Scientific TSX Series Specifications

General IconGeneral
TypeLaboratory Refrigerator
Cooling TypeForced air cooling
Defrost TypeAutomatic
Door TypeSolid or glass door options
Controller TypeMicroprocessor
Certifications/ComplianceUL, CE
Energy ConsumptionVaries by model; refer to product specification sheet
Voltage115V
Exterior Dimensions (H x W x D)Varies by model; refer to product specification sheet
Interior Dimensions (H x W x D)Varies by model; refer to product specification sheet
WeightVaries by model; refer to product specification sheet
ShelvesAdjustable shelves
Energy EfficiencyEnergy Star certified models available