5-6
5.3.2 Results check
1) Analysis is performed automatically. If calibration has
been performed correctly, the message “Calibration
Verified: OK” appears. (Fig.1) If the error message
“Calibration Verified: No Good” (Fig.2) appears,
check that the model eye is not dirty or damaged,
and perform measurement again. If the error
message “No Good” still appears, contact Tomey
Corporation or your local distributor.
(Fig.1)
(Fig.2)