5-9
5.5 Replacing consumables
5.5.1 Fuses
Disconnect the power cord from the outlet when
replacing fuses. Otherwise, electric shock may
occur, resulting in death or serious injuries.
Use fuses of the type indicated on the side of the
fuse holder. Always replace two fuses together.
Using another type of fuse may result in failure or
fire.
When the instrument does not work correctly
after fuses have been replaced, there may be
other causes of the problem. Turn off the
instrument immediately and contact your local
distributor.
1) Turn power off and disconnect the power plug from the
outlet.
2) Disconnect the connector of the power cord from the
outlet.
3) Insert a flat screwdriver into the groove of the fuse
holder (1) under the power terminal and remove the
holder case.
4) Replace the blown fuse with a new one.
5) Reverse the procedure above to install the fuse
holder.
5.5.2 Chin rest paper
1) Remove the two chin rest paper pins.
2) Place new chin rest paper on the chin rest and secure
the paper with the paper pins.
(Fig. 1)
(Fig. 1)