8-3
■
8.5 Declaration of Conformity to EMC
When using specified accessories or cables excluding internal parts
sold as repair parts of TOMEY Corporation, electromagnetic emission
of the CASIA2 may increase or electromagnetic immunity may
decrease.
When using handheld or portable communication equipment, locate it
more than 30cm(12inches) away from the CASIA2. It may degrade the
performance of the CASIA2 if the equipment is located closer.
Do not use the CASIA2 adjacent to or stacked on other equipment. If it
needs to be adjacent to or placed on other equipment, it must be
confirmed that CASIA2 operates correctly in such location.
The CASIA2 falls under Group 1, Class B according to EN 55011(CISPR
11). This means that the CASIA2 does not generate RF energy
intentionally in the form of electromagnetic irradiation, inductive
coupling and/or capacitive coupling for processing of materials or
inspection/analysis, and that it is suitable for use in facilities directly
connected to a low-voltage power network that supplies power to
general household facilities and buildings used for home use.
The CASIA2 can be used in a specialized health care facility such as
hospitals and clinics (except anywhere near the source of a strong
EMI) and a general commercial facility such as eye glasses stores.
The CASIA2 requires special attention for EMC and needs to be
installed, serviced and used based on the following information.
Do not use cables other than those provided or specified by TOMEY
Corporation.
The person who connects additional equipment to the signal I/O
section as part of the medical system, shall take responsibility for
ensuring the system conforms to IEC/EN 60601-1-2 requirements.