36
Use Tolerance Mode
To use the tolerance mode as the process shown
below:
1. Select the desired primary measurement
mode by pressing L/C/R/Z button.
2. Configure the proper test frequency and
series/parallel equivalent mode.
3. Perform the operation of CLEAR appropriately
if necessary.
4. Test standard implements or components with
accurate and reliable measured value.
5. Once the desired measured reading is
displayed, press the TOL button once to store
the reading as the nominal value. At this point,
the “TOL” will be displayed on the screen,
indicating that the tolerance mode is activated.
A percentage mode will be shown in the
secondary mode to indicate the percentage
deviation.
NOTE: Before the press of TOL button, the
primary parameter indicated on LCD in any
mode can be taken as the nominal value,
including data hold, MAX, MIN, AVG data
recording, etc.
6. If sorting is not necessary, you can skip this
step. If it is necessary, by pressing the TOL
button you can select the range of 1%, 5%,