Quick Clear Procedure
Below is an example of steps to do open or short clear:
1 Select the primary and secondary function
mode for measurement;
2 Select test frequency;
3 Select equivalent mode;
4 Keep the test terminal open to perform open
clear;
5 Short the test terminal to perform short clear;
6 Connect DUT to start testing after clear.
NOTE:
1. Clear data is just temporarily stored in RAM,
which means that the data will loss after
power-off. Therefore, “clear, prior to
measurement, should be first done after
power-on.
2. Clear data is stored under different frequency,
thus they will be still valid at the change of test
frequency (for instance, under the frequency
of 1kHz, the meter has been cleared; and
when the frequency is back to 1kHz, there is
no need to clear again.)
3. Clear is not concerned with test parameters
and series/parallel equivalent mode. In