SCPI Commands
FREQuency Subsystem
FREQuency <value>
Description: Set measurement frequency
Parameters are 100, 120, 1000, 10000,100000
or 100hz,120hz,1khz,10khz,100khz (model
supported)
Example: FREQuency 100hz
Set 100Hz frequency
FREQuency?
Description: Query the measurement frequency
Return: < 100hz120hz, 1khz, 10khz100khz>
FUNCtion subsystem
FUNCtion:impa < L | C | R | Z >
Description: Select primary parameter
Example: FUNCtion:impa L
Selects L as primary parameter
FUNCtion:impa?
Description: Query primary parameter
Return: <L, C, R, Z, NULL >
FUNCtion:impb < D | Q | THETA | ESR >
Description: Select secondary parameter
Example: FUNCtion:impb D
Select D as secondary parameter