EasyManuals Logo
Home>Topcon>Measuring Instruments>GPT-8001A

Topcon GPT-8001A User Manual

Topcon GPT-8001A
145 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #106 background imageLoading...
Page #106 background image
9 CHECK AND ADJUSTMENT
9-3
11) Calculate the average value of 6) and 10).
[Example]
6) ••••••••••• • 0° 00’ 00”
10) ••••••••••• • 0° 08’ 20”
___________________________
Average value 0° 04’ 10”
12) Climate to the center of prism. Compare the reading horizontal angle value and calculated
value in
11) If the difference is within 1’30”, no problem for use.
Vertical direction confirmation (Do not move Horizontal direction).
13) Carry out as Horizontal direction confirmation.
Compare the reading Vertical angle value and calculated value.
If the difference is within 1’ 30”, no problem for use. If the difference is more than mentioned
value, contact with your Topcon dealer or Topcon.
[Example] 90¡12 30 90¡ 04 30
Average 90¡08 30
Reading to
prism center 90¡08 50
Difference 20

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Topcon GPT-8001A and is the answer not in the manual?

Topcon GPT-8001A Specifications

General IconGeneral
BrandTopcon
ModelGPT-8001A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals