WhisperSoft Muffler
Follow these steps to clean the accessories.
1. Fully immerse the accessory in the cleaning solution.
DO NOT immerse the Transcend Micro device.
NOTE: The HME Cartridge must be removed prior to
cleaning the HME Adapter. The HME cartridge
cannot be cleaned. Discard it after 7 days of
use.
2. While immersed, thoroughly wipe the surface with a
lint-free cotton cloth. Apply firm pressure and ensure
contact with all accessible connection surfaces to
adequately remove soil buildup.
3. Clean the inside of the accessory by lifting then
lowering the ends of the accessory, while the
accessory is filled with cleaning solution.
4. Rinse accessory by immersing in distilled water. Move
the accessory in a back-and-forth motion for
approximately 10 seconds to remove cleaning agent
residue.
5. Dry the outside of the accessory with a dry, lint-free
cotton cloth. Allow the accessory to air dry until the
inside is dry. The length of drying time will depend on
ambient conditions.