25
Warnings
• Do not allow water to enter this device. Transcend Micro
should not be exposed to environmental conditions
where the system may get wet.
• This device is not intended for life support.
• The Transcend Micro must be set up and adjusted by a
trained provider before being used for therapy ramp
and pressure.
• The air temperature produced by this device can be as
much as 10ºF higher than the temperature of the room.
Exercise caution if the room temperature is warmer than
90ºF (32ºC).
• Do not block or otherwise obstruct the exhalation ports
of the mask. Follow the manufacturer’s instructions,
included with your mask.
• This equipment is not suitable for use with oxygen or in
the presence of a flammable anesthetic mixture with air
or oxygen, or with nitrous oxide. Sources of oxygen
must be located more than 1 meter from the equipment
to avoid the risk of fire and burns.
• The Transcend Micro is only to be used with the
supplied or recommended accessories. Use of
accessories not recommended may result in increased
electromagnetic emissions or decreased
electromagnetic immunity of the PAP system and may
be potentially unsafe.
• The Transcend Micro is not defibrillation proof.
• Do not attempt to sterilize Transcend Micro.