7 Technical Data
7.1 Technical Specications
Measure-
ment
technol-
ogy
Light source LED
Detector Photodiode
Measurement principle Fluorescence
Parameters see parameter list, chapter 7.2
Measurement range 0...200 µg/L or 0...500 µg/L
Measurement accuracy ± 5 %
Turbidity compensation no
Data logger no
Reaction time T100 6 s
Measurement interval 3 s
Housing material Stainless steel (1.4571/1.4404) or titanium (3.7035)
Dimensions (L x Ø) 171 mm x 36 mm
Weight
stainless steel 0,5 kg
titanium 0,4 kg
Interface digital
Ethernet (TCP/IP)
RS-232 or RS-485 (Modbus RTU)
Power
con-
sumption
typical < 1 W
with network < 1.6 W
Power supply 12...24 VDC (± 10 %)
Required supervision typically ≤ 0,5 hours per month
Calibration/maintenance
interval
24 months
System compatibility Modbus RTU
Warranty 1 year (EU: 2 years)
INSTALLATION
Max.
pressure
with Subconn 30 bars
with xed cable
3 bars
in ow cell
1 bar, 2...4 L/min
Protection type IP68
Sample temperature +2...+40 °C
Ambient temperature +2...+40 °C
Storage temperature -20...+80 °C
Inow velocity 0.1...10 m/s