The subitem “Cleaning” (menu item “Options”) allows the user to set the cleaning intervals.
To make use of an automatic cleaning interval, this must rst be activated by checking the box and setting a
cleaning interval. The cleaning time and the pause before the measurement can also be dened here.
Cleaning can also be triggered and dened whether the valve and the relay are activated during cleaning.
When the valve switches, the air must come out of the outlet and stop completely
when the switching procedure has nished. If air is not released or if the valve
does not close correctly at the end of the switching procedure, both hoses must
be swapped.
NOTICE
The compressed air valve must never be exposed to pressure greater than 0.7 Mpa
(7 bars). Otherwise, the valve can be damaged.
NOTICE
To ensure that settings made are not lost when reconguring, it is essential to save a restore
point beforehand.
Please note that the measurement and cleaning intervals should be matched to each other
to avoid suspension of measurements. If this is necessary, use the recommended settings
described in the table below.