MANUAL STORE
After completing a measurement and before taking another one,
sample data can be stored manually under the active project. Refer
to pages 8–3 through 8–5 for instructions on creating and selecting
the active project.
To manually store measurement data, press the 〈STORE〉 key while
data is being displayed. If no project is active when the
〈STORE〉
key is pressed, the gauge displays the error message No active
project!, then returns to the Ready screen.
NOTE
The 〈STORE〉 key is active only when measurement data
is displayed. Measurement data is displayed at the end
of a reading, or the most recent test results can be
displayed using the
〈RECALL〉 key (see page 4–17).
For each measurement, the gauge can store a location description of
up to 12 characters, as well as a note of up to 15 characters. For an
Asphalt mode measurement, the gauge can also store the location
with respect to the centerline (left, right, or neither), and the distance
from the centerline. Follow the prompts to enter location
information and/or a note.