TAKING A DRIFT TEST
To take a new drift test, press 〈1〉 at the Drift Test menu shown on
page
5–7. The gauge prompts the operator to place the gauge on the
reference standard block with the source rod in the SAFE (shielded)
position. Ensure that the gauge is properly positioned and press the
〈ENTER/START〉 key.
NOTE
If it has been less than three hours or more than eight
hours since the last stat test, the gauge displays an error
message. Press
〈ENTER/START〉 to return to the Setup
menu.
The gauge displays the progress of the drift test (five 4-minute
counts). Upon completion of the counts, the gauge displays the
results of the density drift test, including an indication of whether
the test passed or failed:
Press 〈ENTER/START〉 to view the results of the moisture drift
test. Follow the prompts to scroll through the results.