5 Specifications
NOTE: Specifications are subject to change without notice.
5.1 Physical
Dimensions
Interface module: 17.3 cm 10.5 cm 4.5 cm (6.8 in. 4.1 in.
1.8 in.)
High Flow Module: 15 cm 6.7 cm 6.1 cm (5.9 in. 2.7 in. 1.4 in.).
Low Flow Module: 12.7 cm 5.1 cm 2.8 cm (5.0 in. 2.0 in.
1.1 in.).
Flow
connectors
High Flow Module:
Flow inlet: 22-mm female ISO taper.
Flow outlet: 22-mm male ISO taper.
Low Flow Module:
Flow inlet: 3/8-in.
Flow outlet: 3/8-in.
Weight
Approximately 0.7 kg (1.5 lb), interface with high flow module and
cable.
5.2 Environmental
Temperature
Operating: 5 to 40C (41 to 104F). 15 to 80% relative humidity from 5
to 31C decreasing linearly to 15 to 50% relative humidity at 40C.
Storage: -40 to 70C (-40 to 158F) at 10 to 90% relative humidity,
non-condensing
Atmospheric
Pressure
Operating: 57.1 to 106 kPa (8.28 to 15.37 psia).
Storage: 15000 meters
Conditions
Indoor Use
Operating Altitude up to 4000 m (13,000 ft)
Pollution degree I or II
5.3 Power
Battery Life
3 to 6 hours typical
Battery Type
Lithium-Ion SBL160
AC Adapter
12 VDC 1A Minimum