2-10 AeroTrak
®
Handheld and Remote Airborne Particle Counters
5. Remove the display.
6. Use the display from
service kit 9303 Display
(S7000003). Install new
display using the
following steps:
a. Open the display
connector as shown
in step 4 above.
Insert display cable
and close the
connector.
b. Snap the display on
to the three standoffs
(shown in step 2).
c. Connect the keypad
and power on the
unit.
7. Adjust the display
contrast.
C A U T I O N
The electronic circuits within this instrument are susceptible to
electro-static discharge (ESD) damage. Use ESD precautions to
avoid damage.
Use only a table top with a grounded conducting surface.
Wear a grounded, static-discharging wrist strap.