EasyManua.ls Logo

TSI Instruments Aerotrak 9303 - CHAPTER 6 Troubleshooting

TSI Instruments Aerotrak 9303
102 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6-1
CHAPTER 6
Troubleshooting
Symptom
Possible Cause
Corrective Action
Display jittery or not
working (Model 9303)
Contrast too low.
Open unit and adjust potentiometer on
main board to modify display contrast.
See Display Replacement.
Unseated cables.
Inspect display cable for cuts or exposed
wires. Replace with new display module.
See Display Replacement.
Damaged display.
Replace with new display module. See
Display Replacement.
Instrument does not meet
zero count specification
(<1 particle/5 min)
HEPA filter is not connected properly and
room air is leaking into the HEPA filter
assembly.
Check if the HEPA filter has been tightly
connected to the inlet.
Check if rubber O-ring (black) on the inlet
is in place.
Residual particles from previous samples are
shedding off internal parts and into the optics.
Purge instrument by running the
instrument for 10 to 15 minutes before
attempting zero count test.
The filter assembly or inlet is dirty or
contaminated with particles.
Clean the HEPA filter fitting or run an
extended test to clean out particles.
A leak has developed in the aerosol flow path
(Model 9303).
Tighten all tubing connections
Internal optics have become dirty.
Clean optics. See Optics Block Cleaning.
Pump not working
(Model 9303)
Incorrect flow rate.
Use flowmeter to measure flow rate
(factory default 2.83 L/min). Redo flow
calibration if needed.
No input power to pump.
Check voltage on main board J3
supplying power to pump. See Main
Board Removal and Install step 3 for
position of J3. 6 V DC.
Check the crimp on the pump connector.
Bad Pump.
Replace pump. See Pump and Filter
Replacement.

Table of Contents

Other manuals for TSI Instruments Aerotrak 9303

Related product manuals