3-2 AeroTrak
®
Handheld and Remote Airborne Particle Counters
A remote particle counter (Models 7201, 7301, 7310, 7501, and 7510) is
composed of the following parts:
Cle a n Inlet N o z zle aft e r Cali b r a t ion
Swab Selection
Models
Nozzle (new)
Nozzle (old)
Swab
7301 & 7501
7001381
7001277
Wire Cotton
7310
7001508
7001477
Gray
7201 & 7510
7001507
7001283
Green
1. Remove the SS barbed piece
from the Remote Particle
Counter’s “inlet nozzle”.
2. Wipe down its interior surface
with a swab wet with alcohol
and blow it dry with the air
supply nozzle (explained in
further detail in the steps
that follow).