4. Lift the bottom (side
towards pump) of the
optics block assembly
enough to get access to
the silicon tube
connecting the pump to
the optics block. Use a
needle-nose plier and
pull the tube off the
optics block.
C A U T I O N
The electronic circuits within this instrument are susceptible to
electro-static discharge (ESD) damage. Use ESD precautions to
avoid damage.
Use only a table top with a grounded conducting surface.
Wear a grounded, static-discharging wrist strap.