General Maintenance–Model 9303 2-25
15. Adjust the potentiometer until the power meter reads the power listed
below ±0.5 mW at the beam dump.
40 mW if using Melles Griot laser power meter
38 mW if using Newport laser power meter
16. Turn power off and re-assemble beam block to optics assembly.
17. Re-assemble unit and calibrate.
W A R N I N G
Use caution when removing alignment target and placing laser power
meter so that you do not endanger yourself or others from the laser
radiation exiting the optics block.
Adjust potentiometer for changing laser power
Adjust potentiometer for changing laser power