Getting Started
Up and Running Overview
2
Getting Started
Up and Running Overview
1.
Make sure the wet bulb wick is clean. Fill reservoir with distilled water. DO NOT use
ordinary tap water.
See “check wick and fill natural wet bulb”.
2.
Place the QUESTempº 34 & 36 in the work area in a safe location approximately 3.5 feet
off the ground.
3.
Power On the QUESTempº34 & 36 and check the battery.
If the battery voltage displayed during the power-on sequence is less than or equal
to 6.4 volts, replace or recharge the batteries.
4.
Be aware that the sensors require 10 minutes to stabilize to a new environment.
5.
In the main menu, View will be selected (an indicator arrow denotes the selected menu).
Press the I/O Enter key to select.
6.
Press the Run/Stop key to begin storing data.
7.
Use the Up/Down Arrow keys to set the display to the desired items.
Getting started