9 MAINTENANCE INSTRUCTIONS
9.1 Preventive and Scheduled Maintenance
The maintenance operations described in this chapter need to be
followed as indicated to keep the device in good working condition.
The instructions that follow can easily be carried out by un-trained
personnel and do not require a service technician.
Should the need arise, technical assistance or a service technician can be
requested by either calling your dealer or Tuttnauer USA Co.
Warning:
Before carrying out any preventive maintenance operation, ensure that
there is no pressure in the autoclave.
9.1.1 Daily
1. Clean the door gasket with a mild detergent, water and soft
cloth or sponge. The gasket should be clean and smooth
2. It is recommended to run a B & D test (program 6) as the
first cycle each working day.
9.1.2 Weekly
1. Take out the tray holder and trays. Clean the tray holder and
trays with detergent or a non-abrasive stainless steel cleaner
and water, using a cloth or sponge. Rinse the tray holder and
trays immediately with water to avoid staining the metal.
Caution
Do not use steel wool, steel brush or bleach as this can damage
the chamber and trays!
2. Put a few drops of oil on the 2 door pins and door tightening
bolt.
3. Clean the outer parts of the autoclave with a soft cloth.
9.1.3 Periodically
1. Once every month clean and check the safety valve (see
para. 9.6).
2. Once every month drain the generator (this operation shall
be performed by a qualified technician).
3. Clean the strainer once a month (see para. 9.7).
4. Every 6 months or as needed replace the air filter, (see para.
9.3).
5. Every 6 months clean the electronic box with compressed
air, from inside outward (this operation shall be performed
by a qualified technician).
6. Check the door gasket every 12 months and replace it if
required (see para. 9.4).