52
9.7 Cleaning water outlet strainer
Caution!
Before proceeding, Make sure that the electric cord is disconnected and
there is no pressure in the autoclave.
Warnings
1. Do not touch the strainer’s cover, mounted on the left side of the
autoclave, during and shortly after operation. Touching the hot
strainer’s cover may cause severe injuries.
2. If maintenance operation is performed while strainer cover is hot,
use heat resistant gloves to avoid injuries.
1. Open the strainer cap.
2. Remove the strainer element.
3. Rinse the strainer with water, using a brush if necessary.
4. Reinstall the strainer element.
5. Close the strainer cap.
Cap
FIL175-0027
Gasket
GAS082-0008
Strainer element
FIL175-0046
Strainer Housing
FIL175-0051
Water
strainer