OPERATION
&
MAINTENANCE
MANUAL
Electronic Tabletop Autoclaves
Models 1730, 2340, 2540, 3140, 3850, 3870
E, EK, EA & EKA

Questions and Answers

  • P
    patriciafoxSep 12, 2025
    Why are wraps burned after using Tuttnauer 3140EA?
    • L
      Lauren JohnsonSep 12, 2025
      To address wraps coming out burned from your Tuttnauer Laboratory Equipment, check if the water sensor is dirty. Also, verify that the wall outlet voltage is not too high and ensure items are not touching the chamber walls or bottom. Load the autoclave according to the instructions.
  • W
    wbrooksSep 12, 2025
    What does ‘High Temp’ mean on my Tuttnauer 3140EA Laboratory Equipment?
    • B
      brownjamesSep 12, 2025
      If the 'High Temp' message appears during heat up, it may indicate a faulty temperature sensor. If it occurs at other times, the heating elements might be staying on continuously. In either situation, it's recommended to call a technician for assistance with your Tuttnauer Laboratory Equipment.
  • R
    Ryan CookSep 10, 2025
    Why does my Tuttnauer 3140EA display ‘Low Temp’?
    • M
      Matthew VincentSep 12, 2025
      If the 'Low Temp' message appears on your Tuttnauer Laboratory Equipment, refer to the solutions for the Low Water problem. Additionally, check if the sterilization time is set too long, causing the chamber to run dry.
  • G
    Gregory McbrideSep 9, 2025
    What to do if Tuttnauer Laboratory Equipment displays ‘Low Water’?
    • N
      Natasha FergusonSep 9, 2025
      If the 'Low Water' message is displayed on your Tuttnauer Laboratory Equipment, clean the water level electrode inside the vessel. Also, ensure the door is fully closed with the gasket properly seated, and check that the machine is level. Replace the door gasket if there's any steam leakage.
  • S
    Samantha TorresSep 8, 2025
    What to do if the keyboard of my Tuttnauer 3140EA does not respond?
    • D
      Dennis CainSep 8, 2025
      If the keyboard of your Tuttnauer Laboratory Equipment does not respond, press the STOP key for 2 seconds to abort any program that may still be running.
  • M
    Melanie OsborneAug 23, 2025
    What should I do if my Tuttnauer 3140EA Laboratory Equipment displays ‘High Pres’?
    • G
      George SheltonAug 23, 2025
      If your Tuttnauer Laboratory Equipment displays ‘High Pres’, clean the air jet. Ensure the sterilization time and temperature are set correctly; if unsure, use the default settings. Also, make sure the autoclave is not too heavily loaded.
  • C
    Christopher LaraSep 6, 2025
    Why is there a loud noise coming from the reservoir of my Tuttnauer Laboratory Equipment?
    • A
      Anthony GarciaSep 6, 2025
      If there is a loud noise coming from the reservoir of your Tuttnauer Laboratory Equipment, the sterilization temperature might be set too high. The maximum setting is 274ºF (134ºC).
  • S
    sfloresSep 5, 2025
    Why does the exhaust stage take so long when running a cycle on my Tuttnauer 3140EA?
    • R
      Rachel MckaySep 5, 2025
      If you are running a ‘glassware’ program on your Tuttnauer Laboratory Equipment, this is normal. The slow exhaust will take between 15 and 20 minutes.
  • S
    smooreSep 4, 2025
    Why is the drying cycle taking so long in my Tuttnauer 3140EA Laboratory Equipment?
    • B
      Brent CooleySep 4, 2025
      If the drying cycle in your Tuttnauer Laboratory Equipment is taking too long, the autoclave may be overloaded. Also, consider replacing the HEPA filter.
  • W
    Whitney HensonSep 3, 2025
    How to fix a Tuttnauer Laboratory Equipment chamber that is not filling?
    • C
      Cynthia RaySep 3, 2025
      If the chamber of your Tuttnauer Laboratory Equipment is not filling, check the water level in the reservoir. Also, inspect the bottom of the reservoir for any obstruction blocking the pickup tube. Clean the water sensor and try filling the machine manually using the Manual Inlet key.

Summary

General Information and Warranty

Incoming Inspection

Inspect the autoclave's exterior and interior for damage upon arrival. Document any issues.

Technical Data and Specifications

Electrical Data Overview

Table detailing electrical specifications, fuses, and circuit breakers for different models.

Detailed Technical Specifications

Comprehensive table of model specifications including chamber size, water volume, tray dimensions, and weight.

Control Panel and Interface

Control Panel Keyboard Functions

Detailed explanation of the program keys and other operational buttons on the control panel.

Displayed Messages and Safety Measures

Explains common error messages, their causes, and corresponding safety measures.

Sterilization Programs Overview

Program 1: Unwrapped Instruments

Details nominal parameters and operational sequence for sterilizing unwrapped instruments.

Program 2: Wrapped Instruments and Porous Loads

Provides settings and sequence for sterilizing wrapped items and porous loads with drying.

Installation and Setup Instructions

Preparation Before Sterilization

Maintenance Instructions

Preventive and Scheduled Maintenance

Overview of daily, weekly, and periodic maintenance tasks to keep the autoclave in good working condition.

Cleaning the Air Jet

Instructions for cleaning the air jet located in the water reservoir to prevent spore test failures.

Cleaning with Chamber Brite

Step-by-step guide for using Chamber Brite cleaner to descale and clean the autoclave.

Troubleshooting Common Issues

Machine Response and Error Messages

Troubleshooting steps for unresponsive machines and initial error displays like 'Low Heat', 'Low Water', 'Low Pres', 'Low Temp'.

Temperature, Pressure, and Printer Errors

Resolving 'High Temp', 'High Pres' errors, printer malfunctions, and continuous paper feed issues.

Operational Problems: Leaks and Exhaust

Addressing door leaks, long exhaust stages, and blocked water outlet strainers.

Drain, Test Failures, and Load Issues

Troubleshooting clogged drains, failed spore tests, wet or burned items, and instrument rusting.

Performance and Filling Issues

Resolving keyboard unresponsiveness, slow heat-up, noise, long drying times, and chamber filling problems.

Tuttnauer 3140EA Specifications

General IconGeneral
Model3140EA
TypeAutoclave
Chamber MaterialStainless Steel
Power Supply50/60Hz
Operating Pressure2.1 Bar
Temperature Range105°C to 138°C