OPERATION
&
MAINTENANCE
MANUAL
Electronic Tabletop Autoclaves
Models 1730, 2340, 2540, 3140, 3850, 3870
E, EK, EA & EKA

Questions and Answers

  • R
    Ruth MasonSep 2, 2025
    Why is my Tuttnauer 3140EA leaking at the door?
    • R
      Raymond FlowersSep 2, 2025
      If your Tuttnauer Laboratory Equipment is leaking at the door, ensure the door is tightened enough and the door gasket is sealing the chamber. If necessary, replace the door gasket.
  • W
    William HuntSep 1, 2025
    What to do if my Tuttnauer 3140EA Laboratory Equipment is not responding?
    • M
      mmorrisSep 1, 2025
      If your Tuttnauer Laboratory Equipment is not responding, ensure the main switch is in the ‘On’ position. Also, verify the power cord is properly connected to both the machine and the mains. Check the reset button on the cut-out thermostat and make sure the circuit breaker hasn't tripped; lift the circuit breaker lever if it has.
  • K
    Kelly LeeAug 31, 2025
    How to clean a very dirty Tuttnauer Laboratory Equipment chamber?
    • J
      Jennifer DelgadoAug 31, 2025
      To clean a very dirty chamber in your Tuttnauer Laboratory Equipment, use Chamber Brite ™. Very dirty machines may require two cleanings.
  • S
    simsblakeAug 30, 2025
    How to unclog the drain on Tuttnauer 3140EA Laboratory Equipment?
    • S
      sean25Aug 30, 2025
      If the drain of your Tuttnauer Laboratory Equipment is clogged: 1. Open the drain by turning it counterclockwise and place a heavy object over the reservoir cover. 2. Blow compressed air into the drain to force debris back into the reservoir, then clean out the reservoir. 3. Alternatively, disassemble the drain valve, clean it, and blow out the line. Reassemble the valve, ensuring the large 'O' ring is properly seated.
  • J
    Joshua BuckleyAug 29, 2025
    How to fix ‘High Pres’ error on Tuttnauer 3140EA Laboratory Equipment?
    • R
      Rhonda WhiteAug 29, 2025
      If the 'High Pres' message appears on your Tuttnauer Laboratory Equipment, the air jet might be clogged; clean it. It could also mean that the heating elements are continuously on, or the temperature sensor is faulty. In these cases, contact a technician.
  • S
    steven34Aug 28, 2025
    What should I do if my Tuttnauer 3140EA shows ‘Low Pres’?
    • T
      Travis RomanAug 28, 2025
      If the 'Low Pres' message is displayed on your Tuttnauer Laboratory Equipment, clean the water level electrode inside the vessel. Also, ensure the door is fully closed, the door gasket is properly seated, and there is no steam leakage. Check the leveling of the machine.
  • S
    Samuel BaldwinAug 27, 2025
    How to troubleshoot Tuttnauer 3140EA machine not responding?
    • M
      Michael Howard IVAug 27, 2025
      If your Tuttnauer Laboratory Equipment machine isn't responding, verify that the main switch is in the ‘On’ position, the power cord is properly connected to both the machine and the mains, and the circuit breaker hasn't tripped. Also, check the reset button on the cut-out thermostat.
  • A
    andersonshannonAug 26, 2025
    How to clean a very dirty chamber in my Tuttnauer 3140EA?
    • W
      Wanda McgeeAug 27, 2025
      If the chamber of your Tuttnauer Laboratory Equipment is very dirty, use Chamber Brite ™ to clean the autoclave; very dirty machines may require two cleanings.
  • A
    Arthur MurphyAug 25, 2025
    What does ‘Low Heat’ mean on my Tuttnauer 3140EA Laboratory Equipment?
    • I
      ianadkinsAug 26, 2025
      If the message ‘Low Heat’ is displayed during heat up on your Tuttnauer Laboratory Equipment, it indicates a bad temperature sensor. If heating elements are remaining on instead of cycling on and off, call for a technician.
  • K
    Kristina ShermanAug 25, 2025
    What to do when my Tuttnauer 3140EA displays ‘High Temp’?
    • R
      robert00Aug 25, 2025
      If your Tuttnauer Laboratory Equipment displays ‘High Temp’, clean the air jet and check for proper voltage at the wall outlet.

Summary

General Information and Warranty

Incoming Inspection

Inspect the autoclave's exterior and interior for damage upon arrival. Document any issues.

Technical Data and Specifications

Electrical Data Overview

Table detailing electrical specifications, fuses, and circuit breakers for different models.

Detailed Technical Specifications

Comprehensive table of model specifications including chamber size, water volume, tray dimensions, and weight.

Control Panel and Interface

Control Panel Keyboard Functions

Detailed explanation of the program keys and other operational buttons on the control panel.

Displayed Messages and Safety Measures

Explains common error messages, their causes, and corresponding safety measures.

Sterilization Programs Overview

Program 1: Unwrapped Instruments

Details nominal parameters and operational sequence for sterilizing unwrapped instruments.

Program 2: Wrapped Instruments and Porous Loads

Provides settings and sequence for sterilizing wrapped items and porous loads with drying.

Installation and Setup Instructions

Preparation Before Sterilization

Maintenance Instructions

Preventive and Scheduled Maintenance

Overview of daily, weekly, and periodic maintenance tasks to keep the autoclave in good working condition.

Cleaning the Air Jet

Instructions for cleaning the air jet located in the water reservoir to prevent spore test failures.

Cleaning with Chamber Brite

Step-by-step guide for using Chamber Brite cleaner to descale and clean the autoclave.

Troubleshooting Common Issues

Machine Response and Error Messages

Troubleshooting steps for unresponsive machines and initial error displays like 'Low Heat', 'Low Water', 'Low Pres', 'Low Temp'.

Temperature, Pressure, and Printer Errors

Resolving 'High Temp', 'High Pres' errors, printer malfunctions, and continuous paper feed issues.

Operational Problems: Leaks and Exhaust

Addressing door leaks, long exhaust stages, and blocked water outlet strainers.

Drain, Test Failures, and Load Issues

Troubleshooting clogged drains, failed spore tests, wet or burned items, and instrument rusting.

Performance and Filling Issues

Resolving keyboard unresponsiveness, slow heat-up, noise, long drying times, and chamber filling problems.

Tuttnauer 3140EA Specifications

General IconGeneral
Model3140EA
TypeAutoclave
Chamber MaterialStainless Steel
Power Supply50/60Hz
Operating Pressure2.1 Bar
Temperature Range105°C to 138°C