15
2.10.1 Water for the vacuum system and the Drain Cooling
The feed water supplied to the liquid ring vacuum pump must
meet the following requirements:
Hardness: 0.7 - 2 mmol/l.
Water temperature: shall not exceed 59°F (15°C).
Note:
The use of hard water in the vacuum pump may invalidate the
warranty for the vacuum pump, since it can cause blocking of
the rotor, which can damage the pump.
2.11 Environment Emission Information
A. The peak sound level generated by the sterilizer is < 70 / dBA with
background noise of 60 dB.
B. The total heat transmitted by the sterilizer is:
< 100 W / h (341 btu) for 2540 EHS
< 150 W / h (512 btu) for 3870 EHS.
2.12 Symbol Description
Caution! Consult accompanying documents
Caution! Hot surface.
Caution! Hot steam.
Protective earth (Ground)
On-Off

Other manuals for Tuttnauer 3870 EHS