44
12. At the end of the cycle a buzzer rings and the START light switches
OFF. The message “END” is displayed. The exhaust valve is
opened to prevent formation of a vacuum, if the door is not opened
immediately.
In the event of program failure, the exhaust valve is opened, a
continuous sound followed by beeps is output, the FAIL indicator is
lit and the appropriate error message is displayed.
13. Open the door and unload the sterilized material from chamber.
14. At the end of the day switch off the power by means of the rocker
switch on the front panel.
Attention!
The pressure and temperature of the jacket decreases very
slowly when the equipment is turned off.
8.2 Lifting and carrying
Caution:
Before moving the autoclave, Make sure that the electric cord is
disconnected from the power, that all hoses are disconnected from
water supply and drain and there is no pressure in the chamber.
1. Disconnect the power supply cord.
2. Disconnect the water and drain hoses.
3. Drain the water from the reservoir and vessel.
To avoid injuries, lifting and carrying should be done by at least
two people.
Do not drop this device!
8.3 Loading and unloading the Device
8.3.1 Safety
Protective equipment and clothes and other safety instructions
should be implemented in accordance with local and national
regulations and/or rules!
For proper sterilization - Do not overload the chamber. Only
autoclavable products shall be used; for sterilization of unknown
materials or instruments please refer to the instructions of the
manufacturer of these items.
8.3.2 Loading
Correct loading of the autoclave is essential to successful
sterilizing for several reasons. Efficient air removal from the
chamber and the load will permit effective steam penetration and
saturation, and allow proper drainage of condensate.
Additionally, correct loading will reduce damage to packs and
their contents and maximize efficient use of the sterilizer.
For detailed loading instructions, see para. 6 (Preparation before
sterilization)