33
Add Water Message is displayed, the buzzer is activated four times
and the "Water” LED illuminates if the water level
switch indicates no water in the mineral free water
reservoir and the START button is pressed. In this case
the cycle will not begin until water is added and the
START button is pressed.
Ready Message is displayed if the autoclave is ready to begin a
cycle. This will be when the pressure in the jacket has
reached the proper pressure for the required cycle.
Not Ready Message is displayed if the START key has been
pressed before the pressure in the jacket has reached the
proper pressure for the required cycle.
St. By Message is displayed after the autoclave has been turned
on and while the pressure in the jacket is rising to the
pressure required to begin a cycle.
Gen. Error Message is displayed, the buzzer is activated and the
FAIL indicator illuminates when the water pump runs
continuously for more than 10 minutes, attempting to
fill the jacket. To cancel this alarm the main power
switch must be turned off. After turning power back on,
if the GEN.ERROR alarm occurs again after 2½
minutes, a service technician must be called.
Vac Water Message is displayed when the low water level float, in
the feed water reservoir, detects that the water level is
too low.
If a printer is installed on the system, then the printer will print
any ERROR messages that appear on the LCD Display