35
PRINTER OUTPUT DESCRIPTION
Operator :___________ To be filled in manually by operator.
Autoclave:001 Autoclave identifying number.
02/03/2002 09:04:44 Date and time sterilization cycle ended.
Load number: 00011 Load number. Useful for record keeping and scheduling
maintenance
Cycle ended
−−−−−−−−−−−−−−−−−−−
D30:17 187.5°F 26.2Ihg The time, temperature and pressure during post
vacuum drying
-----------------------
E26:15 230.2°F 01.3Psig The time, temperature and pressure during exhaust.
-----------------------
S24:21 276.1°F 31.0Psig The time, temperature and pressure during sterilization.
*
* Prints sterilization data every 1 minute.
*
S18:19 275.5°F 31.9Psig The time, temperature and pressure during sterilization.
S17:19 273.6°F 31.2Psig The time, temperature and pressure during sterilization.
----------------------
H11:25 187.7°F 22.4Ihg The time, temperature and pressure during heating.
----------------------
V09:34 232.4°F 08.8Psig The time, temperature and pressure during pre vacuum.
*
V00:00 076.0°F 00.2Ihg The time, temperature and pressure during pre vacuum.
−−−−−−−−−−−−−−−−−−−
Dry time: 020min Drying time for selected program.
Ster time: 008.0min Sterilization time for selected program.
Ster Temp: 273°F Sterilization temperature for selected program.
PROGRAM: 2-WDry Selected program: 2- 273ºF with dry cycle
02/03/2002 08:17:30 Date and time sterilization cycle began.
EHSEn42-20USA Software version number
Legend
V Vacuum stage
H Heating stage
S Sterilization stage
E Exhaust stage
D Drying stage
Psig Pound per Square Inch
Ihg Inch of Mercury