EasyManua.ls Logo

Tuttnauer 3870 EHS - Page 37

Tuttnauer 3870 EHS
58 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
35
PRINTER OUTPUT DESCRIPTION
Operator :___________ To be filled in manually by operator.
Autoclave:001 Autoclave identifying number.
02/03/2002 09:04:44 Date and time sterilization cycle ended.
Load number: 00011 Load number. Useful for record keeping and scheduling
maintenance
Cycle ended
D30:17 187.5°F 26.2Ihg The time, temperature and pressure during post
vacuum drying
-----------------------
E26:15 230.2°F 01.3Psig The time, temperature and pressure during exhaust.
-----------------------
S24:21 276.1°F 31.0Psig The time, temperature and pressure during sterilization.
*
* Prints sterilization data every 1 minute.
*
S18:19 275.5°F 31.9Psig The time, temperature and pressure during sterilization.
S17:19 273.6°F 31.2Psig The time, temperature and pressure during sterilization.
----------------------
H11:25 187.7°F 22.4Ihg The time, temperature and pressure during heating.
----------------------
V09:34 232.4°F 08.8Psig The time, temperature and pressure during pre vacuum.
*
V00:00 076.0°F 00.2Ihg The time, temperature and pressure during pre vacuum.
Dry time: 020min Drying time for selected program.
Ster time: 008.0min Sterilization time for selected program.
Ster Temp: 273°F Sterilization temperature for selected program.
PROGRAM: 2-WDry Selected program: 2- 273ºF with dry cycle
02/03/2002 08:17:30 Date and time sterilization cycle began.
EHSEn42-20USA Software version number
Legend
V Vacuum stage
H Heating stage
S Sterilization stage
E Exhaust stage
D Drying stage
Psig Pound per Square Inch
Ihg Inch of Mercury

Other manuals for Tuttnauer 3870 EHS

Related product manuals