OPERATION
&
MAINTENANCE
MANUAL
Electronic Tabletop Autoclaves
Models 1730, 2340, 2540, 3140, 3850, 3870
E, EK, EA & EKA

Other manuals for Tuttnauer 3870E

Questions and Answers

  • S
    Shannon ThomasSep 23, 2025
    What to do if Tuttnauer Laboratory Equipment printer doesn't print?
    • B
      Beverly HenrySep 23, 2025
      If the printer on your Tuttnauer Laboratory Equipment isn't printing, first make sure the paper is inserted correctly. If the issue persists, switch off the machine and switch it back on while pressing the feed button on the printer. If a test printout appears, the printer is likely okay, and there may be an electronics issue. Contact your dealer to solve the problem. If no test printout occurs, there is a problem with the printer. Contact your dealer to solve the problem.
  • J
    James SmithSep 23, 2025
    Why does Tuttnauer 3870E exhaust stage take so long?
    • C
      Courtney OsborneSep 23, 2025
      If the exhaust stage of your Tuttnauer Laboratory Equipment cycle is taking a very long time, it's normal if you are running a ‘liquids’ program. The slow exhaust can take between 15 and 20 minutes.
  • P
    powellmaryAug 31, 2025
    What causes 'High Pres' to be displayed on Tuttnauer 3870E?
    • T
      Thomas CoxAug 31, 2025
      The 'High Pres' message on your Tuttnauer Laboratory Equipment can be due to a few reasons. It could mean the air jet is clogged and needs cleaning. Another possibility is that the heating elements are staying on instead of cycling. A faulty or dirty temperature sensor (PT100) could also be the cause. For the latter two issues, it's best to call a technician.
  • L
    lisaallenAug 25, 2025
    What to do when my Tuttnauer 3870E displays ‘High Temp’?
    • D
      Dr. Tracy Bates MDAug 25, 2025
      If your Tuttnauer Laboratory Equipment displays ‘High Temp’, clean the air jet and check for proper voltage at the wall outlet.
  • M
    Michael MeyerAug 25, 2025
    What does ‘Low Heat’ mean on my Tuttnauer 3870E?
    • M
      Mr. Christopher WhiteAug 26, 2025
      If the message ‘Low Heat’ is displayed during heat up on your Tuttnauer Laboratory Equipment, it indicates a bad temperature sensor. If heating elements are remaining on instead of cycling on and off, call for a technician.
  • K
    Kathryn GlassAug 26, 2025
    How to clean a very dirty chamber in my Tuttnauer 3870E Laboratory Equipment?
    • K
      Kathryn FarmerAug 26, 2025
      If the chamber of your Tuttnauer Laboratory Equipment is very dirty, use Chamber Brite ™ to clean the autoclave; very dirty machines may require two cleanings.
  • J
    Joshua PittmanAug 27, 2025
    What to do if Tuttnauer 3870E process fails during heating stage with a HIGH TEMP message?
    • P
      Paul SimpsonAug 27, 2025
      If your Tuttnauer Laboratory Equipment process fails while initiating the heating stage and displays a HIGH TEMP message, try changing the setting of the thermostat.
  • L
    Linda GatesAug 27, 2025
    How to troubleshoot Tuttnauer 3870E Laboratory Equipment machine not responding?
    • C
      Carol YoungAug 27, 2025
      If your Tuttnauer Laboratory Equipment machine isn't responding, verify that the main switch is in the ‘On’ position, the power cord is properly connected to both the machine and the mains, and the circuit breaker hasn't tripped. Also, check the reset button on the cut-out thermostat.
  • D
    David AshleyAug 28, 2025
    What should I do if my Tuttnauer 3870E Laboratory Equipment shows ‘Low Pres’?
    • A
      Anthony StoutAug 28, 2025
      If the 'Low Pres' message is displayed on your Tuttnauer Laboratory Equipment, clean the water level electrode inside the vessel. Also, ensure the door is fully closed, the door gasket is properly seated, and there is no steam leakage. Check the leveling of the machine.
  • J
    Jacqueline ThompsonAug 28, 2025
    How to fix ‘High Pres’ error on Tuttnauer 3870E Laboratory Equipment?
    • T
      Tiffany GilesAug 28, 2025
      If the 'High Pres' message appears on your Tuttnauer Laboratory Equipment, the air jet might be clogged; clean it. It could also mean that the heating elements are continuously on, or the temperature sensor is faulty. In these cases, contact a technician.

Summary

General Autoclave Information

Autoclave Warranty Details

Details the manufacturer's warranty coverage for components and assembly.

Control Panel and Display Functions

Control Panel Key Operations

Details the functions of program and other keys on the control panel.

Displayed Messages and Safety

Lists common messages and associated safety precautions.

Sterilization Programs and Cycles

Program 1: Unwrapped Instruments

Details the settings and sequence for sterilizing unwrapped instruments.

Program 2: Wrapped Instruments

Settings and sequence for sterilizing wrapped instruments and porous loads.

Program 3: Glassware Sterilization

Settings and sequence for sterilizing glassware with slow exhaust.

Installation and Setup Guide

Electrical Installation Safety

Guidelines for safe electrical connection and grounding.

Loading and Unloading Items

Guidelines for safely loading and unloading items for sterilization.

Preparation Before Sterilization

Autoclave Operating Instructions

Initial Setup and Program Selection

Steps for initial setup, filling water, and selecting a sterilization program.

Autoclave Operation Steps

Instructions on starting the autoclave, monitoring stages, and changing parameters.

Maintenance and Care Procedures

Preventive Maintenance Schedule

Overview of daily, weekly, and periodic maintenance tasks.

Air Jet Cleaning

Instructions for cleaning the air jet to prevent sterilization failures.

Chamber Brite Cleaning

Procedure for cleaning the autoclave using Chamber Brite.

Troubleshooting Guide for Operators

Machine Not Responding

Solutions for when the autoclave is unresponsive.

Low Heat Error Troubleshooting

Troubleshooting steps for the 'Low Heat' error message.

Low Water Error Resolution

Steps to resolve the 'Low Water' indicator.

Low Pressure Error Resolution

Solutions for when the 'Low Pressure' message appears.

High Temperature Error

Troubleshooting for the 'High Temperature' error.

High Pressure Error Resolution

Steps to resolve the 'High Pressure' error message.

Tuttnauer 3870E Specifications

General IconGeneral
TypeAutoclave
Model3870E
Number of Trays3
Weight240 lbs
Chamber MaterialStainless Steel
Sterilization Time Range4-60 Minutes
Sterilization Temperature Range121°C
Power Supply230V
Power Consumption1600 W
Dimensions (WxDxH)25 in x 29 in x 21 in

Related product manuals