OPERATION
&
MAINTENANCE
MANUAL
Electronic Tabletop Autoclaves
Models 1730, 2340, 2540, 3140, 3850, 3870
E, EK, EA & EKA

Other manuals for Tuttnauer 3870E

Questions and Answers

  • B
    boylemichaelAug 29, 2025
    How to unclog the drain on Tuttnauer 3870E?
    • Y
      ylopezAug 29, 2025
      If the drain of your Tuttnauer Laboratory Equipment is clogged: 1. Open the drain by turning it counterclockwise and place a heavy object over the reservoir cover. 2. Blow compressed air into the drain to force debris back into the reservoir, then clean out the reservoir. 3. Alternatively, disassemble the drain valve, clean it, and blow out the line. Reassemble the valve, ensuring the large 'O' ring is properly seated.
  • J
    John FreemanAug 30, 2025
    How to clean a very dirty Tuttnauer 3870E Laboratory Equipment chamber?
    • T
      Thomas FrazierAug 30, 2025
      To clean a very dirty chamber in your Tuttnauer Laboratory Equipment, use Chamber Brite ™. Very dirty machines may require two cleanings.
  • N
    Nicholas WhitakerAug 30, 2025
    Why does Tuttnauer 3870E show 'High Temp'?
    • D
      Daniel RossAug 30, 2025
      If the 'High Temp' message appears on your Tuttnauer Laboratory Equipment during heat up, it could indicate a faulty temperature sensor. Alternatively, the heating elements might be staying on continuously instead of cycling on and off. In either situation, it's recommended to call a technician for assistance.
  • M
    Martha CurryAug 25, 2025
    What should I do if my Tuttnauer Laboratory Equipment displays ‘High Pres’?
    • A
      angela66Aug 25, 2025
      If your Tuttnauer Laboratory Equipment displays ‘High Pres’, clean the air jet. Ensure the sterilization time and temperature are set correctly; if unsure, use the default settings. Also, make sure the autoclave is not too heavily loaded.
  • Q
    qjohnsonSep 1, 2025
    How to troubleshoot a clogged drain in Tuttnauer Laboratory Equipment?
    • K
      Kim OchoaSep 1, 2025
      To address a clogged drain in your Tuttnauer Laboratory Equipment, open the drain by turning it counterclockwise. Place a heavy object over the reservoir cover, and blow compressed air into the drain to force debris back into the reservoir, then clean out the reservoir. If needed, disassemble the drain valve by turning it counterclockwise past the stop point, remove the valve, clean it, and blow out the line. Reassemble the valve by turning clockwise, ensuring the large “O” ring is pressed into the groove on the autoclave.
  • C
    Charles PatelSep 1, 2025
    What to do if my Tuttnauer Laboratory Equipment is not responding?
    • A
      Ashley LarsonSep 1, 2025
      If your Tuttnauer Laboratory Equipment is not responding, ensure the main switch is in the ‘On’ position. Also, verify the power cord is properly connected to both the machine and the mains. Check the reset button on the cut-out thermostat and make sure the circuit breaker hasn't tripped; lift the circuit breaker lever if it has.
  • K
    Kristen JonesSep 2, 2025
    Why is my Tuttnauer Laboratory Equipment leaking at the door?
    • G
      gonzalezjimSep 2, 2025
      If your Tuttnauer Laboratory Equipment is leaking at the door, ensure the door is tightened enough and the door gasket is sealing the chamber. If necessary, replace the door gasket.
  • C
    christopherlittleSep 3, 2025
    How to fix a Tuttnauer 3870E chamber that is not filling?
    • J
      Justin HaleSep 3, 2025
      If the chamber of your Tuttnauer Laboratory Equipment is not filling, check the water level in the reservoir. Also, inspect the bottom of the reservoir for any obstruction blocking the pickup tube. Clean the water sensor and try filling the machine manually using the Manual Inlet key.
  • D
    David AriasSep 3, 2025
    Why is the drying cycle taking so long in my Tuttnauer 3870E?
    • S
      stevelewisSep 3, 2025
      If the drying cycle in your Tuttnauer Laboratory Equipment is taking too long, the autoclave may be overloaded. Also, consider replacing the HEPA filter.
  • K
    Kendra MackSep 4, 2025
    Why does the exhaust stage take so long when running a cycle on my Tuttnauer 3870E Laboratory Equipment?
    • B
      Brooke ReidSep 4, 2025
      If you are running a ‘glassware’ program on your Tuttnauer Laboratory Equipment, this is normal. The slow exhaust will take between 15 and 20 minutes.

Summary

General Autoclave Information

Autoclave Warranty Details

Details the manufacturer's warranty coverage for components and assembly.

Control Panel and Display Functions

Control Panel Key Operations

Details the functions of program and other keys on the control panel.

Displayed Messages and Safety

Lists common messages and associated safety precautions.

Sterilization Programs and Cycles

Program 1: Unwrapped Instruments

Details the settings and sequence for sterilizing unwrapped instruments.

Program 2: Wrapped Instruments

Settings and sequence for sterilizing wrapped instruments and porous loads.

Program 3: Glassware Sterilization

Settings and sequence for sterilizing glassware with slow exhaust.

Installation and Setup Guide

Electrical Installation Safety

Guidelines for safe electrical connection and grounding.

Loading and Unloading Items

Guidelines for safely loading and unloading items for sterilization.

Preparation Before Sterilization

Autoclave Operating Instructions

Initial Setup and Program Selection

Steps for initial setup, filling water, and selecting a sterilization program.

Autoclave Operation Steps

Instructions on starting the autoclave, monitoring stages, and changing parameters.

Maintenance and Care Procedures

Preventive Maintenance Schedule

Overview of daily, weekly, and periodic maintenance tasks.

Air Jet Cleaning

Instructions for cleaning the air jet to prevent sterilization failures.

Chamber Brite Cleaning

Procedure for cleaning the autoclave using Chamber Brite.

Troubleshooting Guide for Operators

Machine Not Responding

Solutions for when the autoclave is unresponsive.

Low Heat Error Troubleshooting

Troubleshooting steps for the 'Low Heat' error message.

Low Water Error Resolution

Steps to resolve the 'Low Water' indicator.

Low Pressure Error Resolution

Solutions for when the 'Low Pressure' message appears.

High Temperature Error

Troubleshooting for the 'High Temperature' error.

High Pressure Error Resolution

Steps to resolve the 'High Pressure' error message.

Tuttnauer 3870E Specifications

General IconGeneral
TypeAutoclave
Model3870E
Number of Trays3
Weight240 lbs
Chamber MaterialStainless Steel
Sterilization Time Range4-60 Minutes
Sterilization Temperature Range121°C
Power Supply230V
Power Consumption1600 W
Dimensions (WxDxH)25 in x 29 in x 21 in