OPERATION
&
MAINTENANCE
MANUAL
Electronic Tabletop Autoclaves
Models 1730, 2340, 2540, 3140, 3850, 3870
E, EK, EA & EKA

Other manuals for Tuttnauer 3870EA

Questions and Answers

  • M
    Mary GillAug 25, 2025
    What does ‘Low Heat’ mean on my Tuttnauer Laboratory Equipment?
    • V
      Vincent CampbellAug 26, 2025
      If the message ‘Low Heat’ is displayed during heat up on your Tuttnauer Laboratory Equipment, it indicates a bad temperature sensor. If heating elements are remaining on instead of cycling on and off, call for a technician.
  • J
    jhawkinsAug 26, 2025
    How to clean a very dirty chamber in my Tuttnauer Laboratory Equipment?
    • D
      David LongAug 26, 2025
      If the chamber of your Tuttnauer Laboratory Equipment is very dirty, use Chamber Brite ™ to clean the autoclave; very dirty machines may require two cleanings.
  • J
    Jacob BrownAug 27, 2025
    What to do if Tuttnauer 3870EA Laboratory Equipment process fails during heating stage with a HIGH TEMP message?
    • J
      James NorrisAug 27, 2025
      If your Tuttnauer Laboratory Equipment process fails while initiating the heating stage and displays a HIGH TEMP message, try changing the setting of the thermostat.
  • N
    nicholemarshAug 27, 2025
    How to troubleshoot Tuttnauer 3870EA machine not responding?
    • S
      Sydney GreenAug 28, 2025
      If your Tuttnauer Laboratory Equipment machine isn't responding, verify that the main switch is in the ‘On’ position, the power cord is properly connected to both the machine and the mains, and the circuit breaker hasn't tripped. Also, check the reset button on the cut-out thermostat.
  • R
    robert43Aug 28, 2025
    What should I do if my Tuttnauer Laboratory Equipment shows ‘Low Pres’?
    • A
      Ashley DavidsonAug 28, 2025
      If the 'Low Pres' message is displayed on your Tuttnauer Laboratory Equipment, clean the water level electrode inside the vessel. Also, ensure the door is fully closed, the door gasket is properly seated, and there is no steam leakage. Check the leveling of the machine.
  • B
    Brian NicholsAug 29, 2025
    How to fix ‘High Pres’ error on Tuttnauer 3870EA Laboratory Equipment?
    • D
      Dr. Debra HernandezAug 29, 2025
      If the 'High Pres' message appears on your Tuttnauer Laboratory Equipment, the air jet might be clogged; clean it. It could also mean that the heating elements are continuously on, or the temperature sensor is faulty. In these cases, contact a technician.
  • D
    Derek WilliamsAug 30, 2025
    How to unclog the drain on Tuttnauer 3870EA Laboratory Equipment?
    • M
      Maria HunterAug 30, 2025
      If the drain of your Tuttnauer Laboratory Equipment is clogged: 1. Open the drain by turning it counterclockwise and place a heavy object over the reservoir cover. 2. Blow compressed air into the drain to force debris back into the reservoir, then clean out the reservoir. 3. Alternatively, disassemble the drain valve, clean it, and blow out the line. Reassemble the valve, ensuring the large 'O' ring is properly seated.
  • T
    Travis PetersonAug 23, 2025
    What should I do if my Tuttnauer Laboratory Equipment displays ‘High Pres’?
    • J
      Jose MartinAug 23, 2025
      If your Tuttnauer Laboratory Equipment displays ‘High Pres’, clean the air jet. Ensure the sterilization time and temperature are set correctly; if unsure, use the default settings. Also, make sure the autoclave is not too heavily loaded.
  • C
    Christopher MatthewsSep 1, 2025
    What to do if my Tuttnauer 3870EA is not responding?
    • J
      Jose PayneSep 1, 2025
      If your Tuttnauer Laboratory Equipment is not responding, ensure the main switch is in the ‘On’ position. Also, verify the power cord is properly connected to both the machine and the mains. Check the reset button on the cut-out thermostat and make sure the circuit breaker hasn't tripped; lift the circuit breaker lever if it has.
  • G
    george52Sep 2, 2025
    Why is my Tuttnauer 3870EA Laboratory Equipment leaking at the door?
    • J
      Jessica ChangSep 2, 2025
      If your Tuttnauer Laboratory Equipment is leaking at the door, ensure the door is tightened enough and the door gasket is sealing the chamber. If necessary, replace the door gasket.

Summary

GENERAL AUTOCLAVE INFORMATION

Incoming Inspection Procedure

Inspect the autoclave upon arrival for any shipping damage.

CONTROL PANEL AND INTERFACE

Control Panel Keyboard Functions

Details the functions of each key on the control panel and program selection.

System Messages and Safety Precautions

Details error messages, warnings, and safety precautions displayed by the unit.

STERILIZATION PROGRAMS

Program 1: Unwrapped Instruments

Instructions and parameters for sterilizing unwrapped instruments.

Program 2: Wrapped & Porous Loads

Parameters and procedure for sterilizing wrapped items and porous loads.

Program 3: Glassware Sterilization

Specific program settings for sterilizing glassware.

INSTALLATION INSTRUCTIONS

Electrical Installation and Safety

Guidelines for electrical connection and safety precautions.

Safe Lifting and Carrying

Safety instructions for moving the autoclave.

STERILIZATION PREPARATION

Instrument Cleaning and Preparation

Pre-sterilization cleaning and preparation of instruments.

Sterilization Indicators and Testing

Use of indicators and spore tests for sterilization assurance.

OPERATING INSTRUCTIONS

Initial Autoclave Operation

Steps for initial operation and setup before first use.

MAINTENANCE INSTRUCTIONS

Preventive and Scheduled Maintenance

Overview of routine maintenance tasks to keep the autoclave in good working order.

TROUBLESHOOTING GUIDE

Machine Not Responding

Solutions for when the autoclave does not power on or respond.

Low Heat Error Message

Troubleshooting steps for the 'Low Heat' error message.

Low Water Error Message

Solutions for the 'Low Water' error message.

Low Pressure Error Message

Troubleshooting steps for the 'Low Pressure' error message.

Low Temperature Error Message

Solutions for the 'Low Temperature' error message.

High Temperature Error Message

Troubleshooting steps for the 'High Temperature' error message.

High Pressure Error Message

Solutions for the 'High Pressure' error message.

Door Leakage Problems

Solutions for leaks occurring at the autoclave door.

Clogged Drain Resolution

Steps to clear a clogged drain in the autoclave.

Sterilization Indicator Failures

Troubleshooting issues with sterilization indicators.

Wrapped Items Not Drying

Solutions for wrapped items not drying properly.

Tuttnauer 3870EA Specifications

General IconGeneral
Frequency50/60 Hz
Power Supply120V or 230V
Temperature Range105°C to 137°C
Control SystemMicroprocessor control
Trays Capacity3
Voltage220-240 V
Operating Temperature5°C to 40°C
Chamber Volume85 liters