Tuttnauer 5075 - Service Manual

Tuttnauer 5075
85 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
OPERATION
&
MAINTENANCE
MANUAL
Laboratory Vertical Steam Sterilizers
models 5050, 5075
ELV-D
Standard Autoclave
ELVC-D
Including Fast Cooling System
Cat. No. MAN205-0451001EN Rev B
Tuttnauer Europe b.v., Paardeweide 36, 4824 EH, Breda, P.O. Box 7191, 4800 GD Breda, Netherlands. +31/76-5423510,
Fax: +31/76-5423540
Tuttnauer U.S.A. Co, Ltd. 25 Power Drive Hauppauge, NY, 11788, USA. Tel (631) 737 4850, (800) 624 5836,
Fax: (631) 737 0720

Table of Contents

Other manuals for Tuttnauer 5075

Questions and Answers

  • T
    Terrance WardSep 12, 2025
    Why does my Tuttnauer Steriliser show 'Low Temp' and what should I do?
    • K
      Kristina FlynnSep 12, 2025
      If your Tuttnauer Steriliser displays 'Low Temp', it means the temperature dropped below the sterilization temperature for more than 1 second during the sterilization cycle. To resolve this, perform a new cycle.
  • D
    debbiejonesSep 12, 2025
    What to do if my Tuttnauer Steriliser displays 'Low Pressure'?
    • C
      curtisnicoleSep 13, 2025
      If the Tuttnauer Steriliser displays 'Low Pressure', it means the chamber pressure dropped below the sterilization pressure (134°C = 304 kPa, 121°C = 205 kPa) for 2 seconds during the sterilization stage. To solve this, perform a new cycle.
  • D
    Diane WilsonSep 13, 2025
    What does it mean if my Tuttnauer Steriliser shows 'High Pressure'?
    • S
      Sonya Villegas PhDSep 13, 2025
      If the Tuttnauer Steriliser displays 'High Pressure', it means the Chamber Pressure rose 4.2 psi-29 kPa above sterilization pressure (134°C = 304 kPa, 121°C = 205 kPa) for 2 seconds during the sterilization stage. Perform a new cycle.
  • M
    Megan JordanSep 13, 2025
    What does 'Air Error' mean on my Tuttnauer Steriliser, and how can I fix it?
    • M
      Mark JacksonSep 13, 2025
      If the Tuttnauer Steriliser displays 'Air Error', it means the autoclave didn't reach atmospheric pressure 10 minutes after the drying stage started. Wait until the autoclave reaches atmospheric pressure and then perform a new cycle.
  • T
    Tracy MillsSep 13, 2025
    Why is my Tuttnauer Steriliser not responding?
    • A
      Andrew PaceSep 13, 2025
      If the Tuttnauer Steriliser is not responding, it could be due to several reasons: 1. The main switch might be in the ‘OFF’ position. Turn the main switch to the ‘On’ position. 2. The power cord might be disconnected. Ensure the power cord is properly connected to both the machine and the mains. 3. The circuit breaker might have tripped. Lift the circuit breaker lever.
  • K
    Kyle HillSep 13, 2025
    What does 'Pressure Time Error' mean on my Tuttnauer Steriliser, and how can I troubleshoot it?
    • J
      joseph78Sep 14, 2025
      If the Tuttnauer Steriliser displays 'Pressure Time Error', it means the system couldn't reach the required pressure conditions in the chamber after the preset time during the air removal stage. Verify that the autoclave is not overloaded.
  • K
    Keith BurchSep 14, 2025
    What does the error 'Heat Time Error (Keep)' indicate on my Tuttnauer Steriliser?
    • E
      Erica PerrySep 14, 2025
      If the Tuttnauer Steriliser displays 'Heat Time Error (Keep)', it means the system couldn't reach the required temperature in the chamber during the optional "Keep Heat" stage within the preset time. Verify that the autoclave is not overloaded.
  • M
    Micheal MontgomerySep 14, 2025
    What does 'Low Pressure (Cooling)' mean on my Tuttnauer Steriliser, and how do I fix it?
    • J
      jeffery17Sep 14, 2025
      If the Tuttnauer Steriliser displays 'Low Pressure (Cooling)', it means the pressure in the chamber didn't reach the preset pressure before the cooling stage. Check and fix the compressed air supply.
  • M
    Mr. John Gordon DDSSep 14, 2025
    What does 'High Pressure (Ending)' mean on my Tuttnauer Steriliser, and how do I resolve it?
    • X
      xblanchardSep 14, 2025
      If the Tuttnauer Steriliser displays 'High Pressure (Ending)', it means the system cannot reach atmospheric pressure ± 5kPa during the ending stage. Perform a new cycle.
  • B
    Brian WardSep 14, 2025
    What does 'High Pressure (Exhaust)' indicate on my Tuttnauer Steriliser?
    • R
      Rita BerrySep 15, 2025
      If the Tuttnauer Steriliser displays 'High Pressure (Exhaust)', it means the system cannot reach preset pressure within 10 minutes from the beginning of the exhaust stage. You should perform a new cycle.

Summary

OPERATION & MAINTENANCE MANUAL

Laboratory Vertical Steam Sterilizers 5050, 5075

Details the 5050 and 5075 vertical steam sterilizers.

1. GENERAL

Incoming Inspection and Warranty

Covers initial inspection of the unit and warranty terms.

2. SAFETY INSTRUCTIONS

General Safety Guidelines

Crucial safety precautions for autoclave operation, maintenance, and usage.

3. TECHNICAL DATA

Introduction and Overview

Introduces models and their purpose.

Operating Conditions and Standards

Specifies environmental, operational limits, and compliance standards.

Specifications and Electrical Data

Details technical specifications, dimensions, and electrical requirements.

Utilities and Loading Capacities

Lists required utilities and maximum loading capacities for the unit.

FRONT VIEW

Component Identification

Labels and identifies components shown in the front view diagram.

REAR VIEW

Component Identification

Labels and identifies components shown in the rear view diagram.

OVERALL DIMENSIONS

Dimensional Drawings

Provides visual representations and dimensions of the sterilizer models.

Water Quality Requirements

Built-in Safety Features

Description of Operation Stages

4. CONTROL PANEL

Control Panel Layout

Illustrates the layout of the control panel.

Display Screen Functions

Details the information displayed on the LCD screen during operation.

Keypad Operation Guide

Explains the functions and usage of the control panel keys.

Error Messages and Symbols

Lists and explains error messages and symbols for troubleshooting.

Operational Messages

Describes operational messages and their required user actions.

5. STERILIZATION PROGRAMS

Program Overview and Parameters

Summarizes available sterilization programs and their settings.

Program 1: Unwrapped Instruments

Details parameters and sequence for unwrapped instrument sterilization.

Program 2: Wrapped Instruments

Details parameters and sequence for wrapped instrument sterilization.

Program 3: Unwrapped Pouches

Details parameters and sequence for unwrapped pouch sterilization.

Program 4: Wrapped Pouches

Details parameters and sequence for wrapped pouch sterilization.

Program 5: Unwrapped Delicate Instruments

Details parameters and sequence for unwrapped delicate instrument sterilization.

Program 6: Wrapped Delicate Instruments

Details parameters and sequence for wrapped delicate instrument sterilization.

Program 7: Custom 1

Allows user configuration of parameters for custom sterilization cycles.

Program 8: Custom 2

Allows user configuration of parameters for custom sterilization cycles.

Program 9: Liquid A

Specifies parameters for sterilizing liquids in unsealed bottles.

Program 10: Liquid B

Specifies parameters for sterilizing liquids in sealed bottles.

6. PARAMETERS AND DATA MANAGEMENT

System Navigation Overview

Outlines the system's menu structure for accessing settings and data.

Table of Default Cycle Parameters

Provides a table of default parameters for custom programs.

Accessing the Main Menu

Guides users on how to navigate to the main system menu.

Custom Cycle Parameter Settings

Details how to configure parameters for Custom 1 and Custom 2 programs.

System Parameters Adjustment

Explains how to adjust system-wide settings.

Cycle History Management

Covers viewing and exporting historical cycle data.

12. SERVICE AND MAINTENANCE

Preventive Maintenance Schedule

Details scheduled tasks for maintaining the autoclave's condition.

Daily, Weekly, and Periodic Maintenance

Covers daily, weekly, and periodic cleaning and checks.

Safety Valve Checking Procedure

Procedure for testing the safety valve's functionality.

13. TROUBLESHOOTING GUIDE

Common Malfunctions and Solutions

Provides solutions for common malfunctions and error messages.

14. SPARE PARTS LIST

Available Spare Parts

Lists available spare parts with their catalog numbers.

15. ACCESSORIES

Optional Accessories

Lists optional accessories with their catalog numbers.

Tuttnauer 5075 Specifications

General IconGeneral
TypeAutoclave
Chamber Volume75 liters
Power Supply230V, 50/60Hz
Sterilization Temperature Range121°C to 134°C
Number of Trays3
Chamber MaterialStainless Steel
Weight120 kg
Power3.0 kW
Power Consumption3, 000 watts
Temperature Range105°C to 137°C
Sterilization Time Range5 to 60 minutes
Exterior MaterialSteel
Safety FeaturesOver-temperature protection, pressure relief valve
Cycle TimeVaries depending on load size and sterilization parameters

Related product manuals