The system error is the most general form of error. The
only way to exit from a System Error is to power down
the pump. Turn the pump o and back on to try to clear
the error. If the error cannot be resolved in a timely
manner, power the pump down and contact Technical
Service. The error number displayed on the screen will
reference the type of System Error.
System Error
SYMPTOM PROBABLE CAUSE CORRECTIONS
0 Internal timer stopped.
1 Flash Chip missing OR version incompatible.
2 Too many graphic screens in memory.
4 Never got a ADC ISR for motor current.
5 Never got a ADC ISR for battery level.
6 Never got a ADC ISR for battery temperature.
7 Never got a ADC ISR for Ultrasonic Upstream Occlusion or Bag-
Empty checks.
8 Never got a ADC ISR for Ultrasonic during Downstream
Occlusion.
9 RS232 Communications Transmission Error.
10 Flash Write Error.
11 Flash Read Error.
12 General Stopcock Time Out.
13 Time Out for Buzzer Test
16 Flash Screen Read Error.
19 Flash Languages Read Error.
20 Could not nd Stop Cock Flush Position during prime routine.
21 Could not nd Stop Cock Feed Position during prime routine.
22 Motor Error during Tubing Auto Load.
23 Motor Failsafe Time Out Error. Possible damaged gearbox
encoder or ex cable.
25 Display Lockup Time Out Error.
33 Motor Control Failure.
Language Selection, First Power Up Immediately after powering up the pump for the rst time, the language option screen will appear. Select the language
using the (
) or (
) buttons. When the language is selected, push the fth button to enter the language selection.
After the language is selected the rst time the pump is powered up, the language selection screen will not appear during
future power up operations. Should the language need to be changed after the initial power up, it can be changed through
the “MORE OPTIONS” menu.
Keep or Clear Prior Pump Settings Immediately after powering up the pump, if the prior feed or ush settings were not cleared and the settings are not locked
in the BioTech Screen, the pump will give the following two options:
KEEP SETTINGS Select this option to start with the same settings that were most recently programmed into the pump.
Any of the previously-programmed settings can be modied, if needed. The feed volume totals, “mL FED” (Continuous Feed
Mode) or “BOLUSES FED” (Intermittent Feed Mode) and “mL FLUSH” are also maintained. If a VTBD has been set, the VTBD
setting originally programmed will not change, however, volume to be delivered will start over at one mL.
CLEAR SETTINGS Select this option to reset all input feeding settings and the output feed totals to zero. It will then be
necessary to program all settings before running the pump. The BioTech options settings will not change.
Section VI — Error Codes and Troubleshooting Guide
For operational errors, the KANGAROO ePump enteral feeding pump display will show the error indicator icon, with specic information regarding the error condition, and will
sound the alarm.
The following conditions will trigger error conditions and will activate audible and visible alarms:
Feed or Flush containers are empty
Feed or Flush tube becomes occluded between bag and pump
Tube becomes occluded between pump and patient
Battery low (alarm beeps continuously, about 15 minutes before shut-o)
Unit is left in HOLD mode longer than 10 minutes without input
Pump set tubing is improperly loaded around the rotor
Feeding complete
The MISTIC connector is removed during AUTOPRIMING or RUNNING
• System errors, see Table below
Troubleshooting Guide Listed below are some of the probable causes of alarm conditions and their corrections