ENG3
Section II — Safety and Warnings
CAUTION:  U.S. federal law restricts the sale of this device to physicians or to their direct representatives.
1.  Read this booklet thoroughly before using the KANGAROO ePump enteral feeding pump.
2.  Do not use this device near ammable anesthetics. 
3.  Disconnect power supply before cleaning or servicing.
4.  Use only KANGAROO ePump enteral feeding pump Power Cord (with built-in A/C adapter) with pump.  See Section XII - Assembly Diagrams and Reorder Numbers.
5.  Use only KANGAROO ePump enteral feeding pump Pump Sets with this device.  Pump is not compatible with other Pump Sets.
6.  For re-certication, see Section VII- Re-certication of Performance.  For other integrity checks, consult with a qualied Biomedical technician or contact the manufacturer  
(Section XIII- Customer Service).
7.  For service or for technical information, please contact Customer Service (Section XIII).
8.  Dispose of old battery-powered electronic equipment in a manner consistent with institutional policy for expired equipment disposal.
9.  Cleaning frequency and practices must be consistent with institutional policy for cleaning of non-sterile devices.  See Section IV - Cleaning, for instructions on cleaning the 
KANGAROO ePump enteral feeding pump.
10.  For additional safety information relating to icon descriptions, specications and/or symbols, see Section  III - Icon, Specications, and Conformity. 
11.  This device is designed to minimize the eects of uncontrolled electromagnetic interference and other types of interference from external sources.  Avoid use of other 
equipment that may cause erratic operation or degradation in the performance of this device.
12.  WARNING:  The serial I/O port (see Figure 1A) is intended only to be used with a specically certied accessory to retrieve/extract certain information from the pump. This 
accessory is specically certied as complying with the safety requirements of accessories used with the pump, whereas the safety certication of the accessory has been 
performed in accordance with IEC 60601-1 and as appropriate the IEC/EN 60601-1 harmonized standard. 
  Any accessory equipment that can be connected to the DB9 Serial I/O port, and which does not comply with the equivalent safety requirements of this equipment, may lead 
to a reduced level of safety of the pump.  While use of the pump’s DB9 port is not recommended for customers, the following should be considered in the event that the 
customer decides to connect to this port:  
  The accessory is specically certied as complying with the safety requirements for accessories used with the pump, whereas the safety certication of the accessory has been 
performed in accordance to the appropriate IEC 60601-1 and/or IEC 60601-1-1 harmonized standard. 
13.  CAUTION:  This enteral feeding pump should only be used for patients who can tolerate the ow rates and accuracy levels delivered by the pump.  Premature infants may 
require higher accuracy rates than specied for this enteral feeding pump. For optimal accuracy, the bottom of the formula container should be 18” above the pump, do not 
reuse feeding sets and avoid over stretching the silicone tube that wraps around the pump’s rotor. 
14.  Do not attempt to rotate the valve stem inside the feeding set valve assembly. Rotation of the valve stem can only be done by the pump while the valve assembly is properly 
loaded into the pump.
15.  Should feeding sets require rinsing, it is recommended that the feeding sets be rinsed while they are loaded in the pump. Should the removal of the feeding set valve 
assembly be desired, power the pump o and wait ve seconds before removing. Upon re-insertion of the feeding set valve assembly, make sure the valve assembly is 
properly loaded into the pump.